Level and PressureOperating InstructionsCapacitive electrodes EKProfibus PA
10 Capacitive electrodes EK - Profibus PA321EK 21 up to 6 bartemperatureadapterEK 24:from 100°C 6 bar,max. 150°C for30 min.Product descriptionPEPTFEFE
Capacitive electrodes EK - Profibus PA 111)Ambient temperature on the oscillatorProduct descriptionElectronics temperatureProduct and ambient temperat
12 Capacitive electrodes EK - Profibus PAProduct description1.4 ApprovalsExplosion protectionOnly certified EK** Ex 0 capacitive electrodesmay be used
Capacitive electrodes EK - Profibus PA 13Type EK 31 (partlyinsulated)L (min. 400 mm, max. 20000 mm)Product description1.5 DimensionsTyp EK 21 (fully i
14 Capacitive electrodes EK - Profibus PAProduct descriptionHousing Temperature adapterhousing of plastic/AluminiumConcentric tube Screening tubeof 1.
Capacitive electrodes EK - Profibus PA 15ser. no. 10612892Product description1.6 Type plateBefore mounting and electrical connection,please make sur
16 Capacitive electrodes EK - Profibus PAMountingLateral load2 Mounting2.1 Mounting instructionsGeneralDifferent mediums and measurementrequirements m
Capacitive electrodes EK - Profibus PA 17MountingShortening the electrodeHumidityTo avoid fraying the steel cable (EK 31)during cutting, you must tin
18 Capacitive electrodes EK - Profibus PACable electrod in solidsMountingFixingweightContinuous electrodesLNon-conductive vesselIn non-conductive vess
Capacitive electrodes EK - Profibus PA 19dd d 6 d 6MountingFillingEmptyingMaterial cone, filling and emptying centeredd312 d 10Filling Emptying1 Emp
2 Capacitive electrodes EK - Profibus PASafety information, Note Ex areaSafety informationPlease rfead this manual carefully, and alsotake note of cou
20 Capacitive electrodes EK - Profibus PAElectrical connection3 Electrical connection3.1 Connection instructionsNoteSwitch off the power supply before
Capacitive electrodes EK - Profibus PA 21-+ESCOKTank 1m (d)12.34512 5678PROFIBUS+-EEx ia IICsee certificateon87654321RSENSORDISPLAY1234 56 78Electrica
22 Capacitive electrodes EK - Profibus PAElectrical connectionQualified personnelInstruments which are operated in Ex areasmust only be connected by q
Capacitive electrodes EK - Profibus PA 23Profibus PA in Ex environmentFor application in Ex areas, the PA bus andall connected instruments must be cer
24 Capacitive electrodes EK - Profibus PAElectrical data of the cablesRDCno. of A in Z31.25kHzC in damping screencores mm2nF/kmSINEC 6XV1 44 Ω/km 2 0.
Capacitive electrodes EK - Profibus PA 25Electrical connection+E1-E2+A1+A2-B62-30 WINOUT12 5678PROFIBUS+-EEx ia IICsee certificateon87654321Rcarrier r
26 Capacitive electrodes EK - Profibus PAElectrical connectionWiring example for the installation of an Ex system on vessels without cathodiccorrosion
Capacitive electrodes EK - Profibus PA 27Electrical connectionWiring example for the installation of an Ex system on vessels with cathodiccorrosion pr
28 Capacitive electrodes EK - Profibus PASetup4 Setup4.1 Adjustment mediaAll VEGA Profibus sensors operate in profile3 and can be adjusted with:- the
Capacitive electrodes EK - Profibus PA 29If all sensors are already connected to thebus line, the address must be adjusted viaVEGACONNECT 3, the adjus
Capacitive electrodes EK - Profibus PA 3ContentsContentsSafety information ... 2N
30 Capacitive electrodes EK - Profibus PASetup4.2 Adjustment with the adjustmentmodule MINICOMAs with the PC, the sensor can also beadjusted with the
Capacitive electrodes EK - Profibus PA 31Arrow keys (5)With the keys > and <, you can move withinthe menu level from one menu item to theother.D
32 Capacitive electrodes EK - Profibus PASetupApplication:- highly conductive products- adhesive, conductive products- solids with fluctuating moistur
Capacitive electrodes EK - Profibus PA 33SetupKey DisplayAdjust-mentPara-meteradjustmentSensorm(d) 4.700OKOKwithmediumMin.adjustat %-------(Min. adju
34 Capacitive electrodes EK - Profibus PASetupAdd’lfunc-tionsInfoSensortypeCAPE34SerialnumberSoftw.Vers.SensorTa gmax.rangepFOffsetpFSensoraddr.Sensor
Capacitive electrodes EK - Profibus PA 35SetupMediumconductive Yesact.dist. m X,XXBolt print figures are sensor ormeasured value information andcann
36 Capacitive electrodes EK - Profibus PASensor Sensor Sensor SensorVVOSegmentcouplerProfibus PAcableProfibus DPinterface card(Softing)DP cableVVO con
Capacitive electrodes EK - Profibus PA 37First of all, click to "Profibus DP for PAsensors". Then choose the Profibus cardwhich is installed
38 Capacitive electrodes EK - Profibus PAProfibus PA addressIn this window, you can assign a busaddress to the sensor in the field "Sensoraddress
Capacitive electrodes EK - Profibus PA 39Measurement loop descriptionIn this field, you can give your measurementloop a descriptive label, e.g. levelm
4 Capacitive electrodes EK - Profibus PAProduct description1 Product description1.1 Function and configurationCapacitive electrodes series EK can dete
40 Capacitive electrodes EK - Profibus PAFrom the window "Instrument data parameteradjustment", you can reach all submenus ofthe sensor. Jus
Capacitive electrodes EK - Profibus PA 41SetupMax. adjustmentThere must be max. level in the vessel. If youpush the button Save, the output is set to
42 Capacitive electrodes EK - Profibus PASetupIf you have chosen pF, you can assign acorresponding value in pF to the percentagevalue.Example0 % = 97,
Capacitive electrodes EK - Profibus PA 43Sensor paramter adjustment - OutputsIn this window, you can change the presetvalues for the PA output and the
44 Capacitive electrodes EK - Profibus PASensor parameter adjustment - SensoroptimisationIn this window, the mode of the sensor canbe set. With this,
Capacitive electrodes EK - Profibus PA 45Confirm the question with OK, if theconditions are met.Display Diagnostics Instrument data Configuration Serv
46 Capacitive electrodes EK - Profibus PADiagnosticsDisplay Diagnostics Instrument data Configuration Services Quit HelpStatusSimulationUnder the menu
Capacitive electrodes EK - Profibus PA 47Under the menu item "Services" you canselect the following functions:• Print•View• Backup• Restore
48 Capacitive electrodes EK - Profibus PASetupOperatorThis user level is for the operator. In thislevel, you can have the measured valuedisplayed and
Capacitive electrodes EK - Profibus PA 49Diagnostics5 Diagnostics5.1 SimulationTo simulate a certain filling level, you can callup the function Simula
Capacitive electrodes EK - Profibus PA 5Product description1.2 Types and versionsType 1)EK EK EK EKVersion 11 21 24 31Continuous ••••Partly insulated
50 Capacitive electrodes EK - Profibus PADiagnostics5.4 Failure rectificationFailure Measure, failure rectificationSensor defective, Test the internal
Capacitive electrodes EK - Profibus PA 51DiagnosticsFailure Measure, failure rectificationContact 4 and vesselThe connection between contact 4 and the
52 Capacitive electrodes EK - Profibus PANotesFailure Measure, failure rectificationError messageError messageError messageError messageError message
Capacitive electrodes EK - Profibus PA 53Notes
54 Capacitive electrodes EK - Profibus PANotes
Capacitive electrodes EK - Profibus PA 55Notes
ISO 9001All statements relating to scope of delivery, application, practical useand operating conditions of the sensors and processing systemscorrespo
6 Capacitive electrodes EK - Profibus PAProduct descriptionType 1)EK EK EK EKVersion 11 21 24 31Screening tube (option) stst • • •Temperature adapter(
Capacitive electrodes EK - Profibus PA 7Product description1.3 Technical dataHousingHousing material Plastic PBT (Polyester) or Aluminiumplastic coate
8 Capacitive electrodes EK - Profibus PAProduct descriptionAccessoryFixing spring of 1.4571- length approx. 185 mm (strained)- tensile force approx.
Capacitive electrodes EK - Profibus PA 9+-0 %5 %10 %15 % vol.Product descriptionfully insulatedelectrodepartlyinsulatedelectrodesolidhumidity(%)Level