pLevel and PressureOperating InstructionsHydrostatic pressuretransmitters D80

Seite 2 - Note Ex area

10 Pressure transmitter D80WeightsBasic weight without housing approx. 1.6 kgExternal housing approx. 400 gAdjustment and display elementsPressure tra

Seite 3 - Pressure transmitter D80 3

Pressure transmitter D80 11Product description3005007001824 30 36100Load in ΩSupply voltagein VSupply and signal circuit (analogue transmission, 4 … 2

Seite 4 - 1 Product description

12 Pressure transmitter D80Protective measures 1)Protection IP 66, IP 67, IP 68VEGABOX 01 IP 66 and IP 67Protection class IIIOvervoltage category IIIA

Seite 5 - Pressure transmitter D80 5

Pressure transmitter D80 13Product descriptionOperating conditionsAmbient conditionsAmbient temperature -40°C … +85°C- with indicating module -10°C …

Seite 6 - Electronics version I:

14 Pressure transmitter D80Process fittings D801.7 DimensionsHousingProduct descriptionø 7667ø 7591ø 41,64663ø 41,693ø 41,64690*) VBUS versionø 41,6ø

Seite 7 - , VVO adjustable

Pressure transmitter D80 15Product description8510811813538 72ø 5M20x1,513013972100with protective coverVEGABOX 01

Seite 8

16 Pressure transmitter D80Mounting2 Mounting2.1 Mounting instructionsThe pressure transmitter can be mounted inany position. Cable entries must point

Seite 9 - 1.5 Technical data

Pressure transmitter D80 17Electrical connection3 Electrical connection3.1 Connection instructionsThe electronics in the pressure transmitterrequires

Seite 10 - 18 24 30 36

18 Pressure transmitter D80Electronics version K4 … 20 mA output signal with integratedadjustment and remote parameter settingthrough PC and HART® han

Seite 11 - Pressure transmitter D80 11

Pressure transmitter D80 19Electrical connection3.3 Connection to external connection housing VEGABOX 01Zum Anschluss anDruckmessumformer mitanalogem

Seite 12 - 12 Pressure transmitter D80

2 Pressure transmitter D80ContentsContentsSafety information ... 2Note Ex area ..

Seite 13 - Operating conditions

20 Pressure transmitter D80Electrical connectionSupply through a VEGA signal conditioning instrumentStandard wiring for non-standardised output.Supply

Seite 14 - 1.7 Dimensions

Pressure transmitter D80 21Setup4 Setup4.1 Adjustment structureThe hydrostatic pressure transmitters comewith or without adjustment capability.After c

Seite 15 - VEGABOX 01

22 Pressure transmitter D80Adjustment with PC and adjustment pro-gram VVO (electronics version B, I, K andL)With the adjustment program VVO (VEGAVisua

Seite 16 - 2 Mounting

Pressure transmitter D80 23SetupAdjustmentFor adjustment of zero and span, an amme-ter must be connected to terminals 1 and 3.The measured value is id

Seite 17 - 3 Electrical connection

24 Pressure transmitter D80 250 Ω Ri < 250 Ω +-PLCSetup4.4 Setup with HART® handheldSensors with electronics versions I, K and Lcan be set up with

Seite 18 - Klemmeinsatz

Pressure transmitter D80 25SetupThe most important adjustment stepsOn the following four pages you see a menuschematic of the HART® handheld in conjun

Seite 19

26 Pressure transmitter D80Setup123451.1 4.1(5.1) 4.2(5.2)Less importantmenu windowsNot required, unimportantor blocked menu windowsImportant andrequi

Seite 20 - + -

Pressure transmitter D80 27SetupBlocked menu. Thefollowing menu window isindicated but not supportedby the sensor. Thereforesaving of adjustmentscarri

Seite 21 - 4.1 Adjustment structure

28 Pressure transmitter D80Setup1.3.31.3.46 Write protect None7 Descriptor8 Message9 PV Snsr s/nFinal asmbly num 0Revision #’s1.4like menu 1.3.41.2.3.

Seite 22 - 4.2 Sensor without adjustment

Pressure transmitter D80 29Setup1.2.3.1.11.2.3.1.2Note:After adjustment of a parameter, push "ENTER" and then "SEND".Confirm the m

Seite 23 - Integration time

Pressure transmitter D80 33 Electrical connection3.1 Connection instructions ... 173.2 Terminal assi

Seite 24 - handheld

30 Pressure transmitter D802221) Ri < 250 Ω 4 ... 20 mA Setup5.5 Adjustment with PC directly on the sensor(electronics versions I, K and L)For conn

Seite 25 - Pressure transmitter D80 25

Pressure transmitter D80 31SetupIf you have already connected the PC withthe adjustment software VVO to your measur-ing system,• first switch on the p

Seite 26 - menu schematic

32 Pressure transmitter D80• Confirm the adjustment with " " and aftersaving you are again in the initial menu.SetupNote:When connecting the

Seite 27 - Pressure transmitter D80 27

Pressure transmitter D80 33You can carry out the min./max. adjustmentwith or without medium. Generally you willcarry out the adjustment without medium

Seite 28 - 28 Pressure transmitter D80

34 Pressure transmitter D80• Click to " ".• When the sensor is unpressurised and ininstallation position, confirm the safetyenquiry with &qu

Seite 29 - Pressure transmitter D80 29

Pressure transmitter D80 35You can choose four linearisation curves:linear, horiz. cylindrical tank, spherical tankand user programmable curve.Under &

Seite 30

36 Pressure transmitter D80Max. 32 index markers (value pairs) can beentered.• Quit the menu with " ".• Confirm the message with " &quo

Seite 31 - User access

Pressure transmitter D80 37Saved sensors can be transferred later on toother sensors.For example, if you have a system with sev-eral of the same stora

Seite 32 - Configuration/Measuring

38 Pressure transmitter D80Diagnostics6 DiagnosticsMaintenanceThe pressure transmitter D80 is maintenancefree.Error rectificationThrough continuous se

Seite 33 - Pressure transmitter D80 33

Pressure transmitter D80 39Instrument modification7 Instrument modification7.1 Retrofitting the adjustment insertSuch a retrofitting can be necessary,

Seite 34 - Parameter adjustment

4 Pressure transmitter D80Product description1 Product description1.1 Function and configurationThe pressure transmitter D80 is an efficientinstrument

Seite 35 - Offset correction

ISO 9001All statements concerning scope of delivery, application, practicaluse and operating conditions of the sensors and processing sys-tems corresp

Seite 36 - 36 Pressure transmitter D80

Pressure transmitter D80 5Product description1.2 Electronics version without adjustmentElectronics version B:Pressure transmitter for connection to VE

Seite 37 - Meas. data info

6 Pressure transmitter D80Product descriptionElectronics version I:Pressure transmitter 4 … 20 mA, HART®, VVOVVO512344-20mAterminalKlemmeinsatz+-12V..

Seite 38 - 1 2 3

Pressure transmitter D80 71.3 Electronics version with integrated adjustment in connection housingElectronics version K:Pressure transmitter 4 … 20 mA

Seite 39 - 7 Instrument modification

8 Pressure transmitter D801.4 Electronics for connection to VEGADIS 12VEGADIS 12DISPLAY+ -TRANSMITTER5876321101112OPERATEPOINTDISPLAYTRANSMITTERENDZE

Seite 40 - ISO 9001

Pressure transmitter D80 9Product description1.5 Technical dataMechanical dataMaterials, wetted partsProcess connection stainless steel 1.4571, 1.4435