Operating InstructionsPressure transmitter with METEC® measuring cellVEGABAR 55Probus PA - climate compensatedDocument ID: 39442
103 Product descriptionVEGABAR 55 • Probus PA - climate compensated39442-EN-1303213.3 OperationThe instrument can be adjusted with the following adju
113 Product descriptionVEGABAR 55 • Probus PA - climate compensated39442-EN-130321parameter adjustment of these instruments, an adjustment software s
124 MountingVEGABAR 55 • Probus PA - climate compensated39442-EN-1303214 Mounting4.1 General instructionsMake sure that all parts of the instrument
134 MountingVEGABAR 55 • Probus PA - climate compensated39442-EN-13032121212112Fig. 4: Position of the lter element1 Filter element2 Blind plugCau
144 MountingVEGABAR 55 • Probus PA - climate compensated39442-EN-1303214.2 Mounting stepsFor mounting VEGABAR 55, a welded socket is required. You c
155 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-1303215 Connecting to power supply5.1 Preparing the connectionAl
165 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-130321next distributor must be connected to each other and also co
175 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Fig. 7: Connection steps 6 and 7Proceed as follows:1. Loose
185 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Information:The cable gland can be mounted in three position
195 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-130321I2CDisplay112 5 678Fig. 10: Wiring plan, single chamber hous
2ContentsVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Contents1 About this document1.1 Function ...
205 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-130321123Display12I2CFig. 12: Connection compartment, double chamb
215 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-130321132Display12 5678I2CFig. 15: Electronics compartment, double
225 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-130321112Fig. 17: Wiring plan, Ex-d double chamber housing1 Volta
235 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-130321567812DisplayI²C21543Fig. 20: Electronics and connection com
245 Connecting to power supplyVEGABAR 55 • Probus PA - climate compensated39442-EN-1303211234634125Fig. 21: Connection of the sensor in the housing s
256 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-1303216 Set up with the display and a
266 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Fig. 23: Insert display and adju
276 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321 – Save value• [->] key to se
286 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321The display and adjustment modul
296 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-1303215. Select the requested unit, e
3ContentsVEGABAR 55 • Probus PA - climate compensated39442-EN-1303218 Maintenanceandfaultrectication8.1 Maintain ...
306 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Information:For an adjustment wi
316 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321You can nd the data on the type
326 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-1303214. Activate the selection with
336 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321If the adjustment ranges are exc
346 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321This function enables reading ou
356 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Menu section Function Resetvalu
366 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Basic adjustment▶Basic adjustmen
376 Set up with the display and adjustment module PLICSCOMVEGABAR 55 • Probus PA - climate compensated39442-EN-130321ServiceBasic adjustment 4Display
387 Set up with PACTware and other adjustment programsVEGABAR 55 • Probus PA - climate compensated39442-EN-1303217 Set up with PACTware and other ad
397 Set up with PACTware and other adjustment programsVEGABAR 55 • Probus PA - climate compensated39442-EN-1303217.2 Parameter adjustment with PACTw
41 About this documentVEGABAR 55 • Probus PA - climate compensated39442-EN-1303211 About this document1.1 FunctionThis operating instructions manual
408 Maintenance and fault recticationVEGABAR 55 • Probus PA - climate compensated39442-EN-1303218 Maintenanceandfaultrectication8.1 MaintainIf
418 Maintenance and fault recticationVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Error Cause RecticationWhen an addi-tional instrume
428 Maintenance and fault recticationVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Error code Cause RecticationE113 Communication con-
438 Maintenance and fault recticationVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Total deviation digital output signal absolute:Ftota
448 Maintenance and fault recticationVEGABAR 55 • Probus PA - climate compensated39442-EN-130321At "www.vega.com/downloads" go to "So
459 DismountingVEGABAR 55 • Probus PA - climate compensated39442-EN-1303219 Dismounting9.1 Dismounting stepsWarning:Before dismounting, be aware of
4610 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-13032110 Supplement10.1 Technical dataGeneral dataPressure type Gauge pressure,
4710 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321 Ʋ Connection cable with IP 68 1 bar versionPE, PURWeight approx.0.8 … 8 kg
4810 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321The specications are only an overview and refer to the measuring cell. Lim
4910 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Nominalrange Overload capacity, max. pressureOverload capacity, min. pres-
52 For your safetyVEGABAR 55 • Probus PA - climate compensated39442-EN-1303212 For your safety2.1 Authorised personnelAll operations described in t
5010 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Nominalmeasuringrangeinbar/kPaNominalmeasuring range in psigAverage te
5110 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Long-term drift of the zero signal: Ʋ For one year < 0.05 % x TD Ʋ For
5210 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321 Ʋ Single chamber housing – 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),
5310 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Electromechanical data - version IP 68Connection cable between IP 68 instru
5410 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Electrical protective measuresProtection rating Ʋ Housing, standard IP 66/I
5510 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321bar% Lin%Min-MaxadjustmentLinearizationPROFIBUS PA-outputPrimaryValueSecond
5610 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321ExamplesoftelegramcongurationIn the following you will see how the modu
5710 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321Byte nByte n+1Bit7VZExponentBit62Bit52Bit42Bit32Bit22Bit12Bit02Bit72Bit62Bi
5810 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-13032110.3 DimensionsPlastic housing~ 69 mm(2.72")ø 79 mm(3.03")112 mm
5910 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321AluminiumhousinginprotectionratingIP66/IP68(1bar)120 mm (4.72"
62 For your safetyVEGABAR 55 • Probus PA - climate compensated39442-EN-1303212.6 CE conformityThis device fullls the legal requirements of the appl
6010 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321StainlesssteelhousinginprotectionratingIP66/IP68(1bar)117 mm (4.6
6110 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-13032112368 mm(2.68")68 mm(2.68")92 mm(3.62")42mm(1.65")40mm(
6210 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321VEGABAR55,threadedttingGG GL - 180°CGNGG - 180°C/200°CBA / BB BE78 mm(3
6310 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321VEGABAR55,hygienictting178 mm(3 5/64")78 mm(3 5/64")ø 64 mm(
6410 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321VEGABAR55,hygienictting278 mm(3 5/64")78 mm(3 5/64")78 mm(3
6510 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-130321VEGABAR55,angeconnection3 15/32"4 1/64"5 7/16" 1/8"
6610 SupplementVEGABAR 55 • Probus PA - climate compensated39442-EN-13032110.4 Industrial property rightsVEGA product lines are global protected by
67INDEXVEGABAR 55 • Probus PA - climate compensated39442-EN-130321INDEXAAccessories – Display and adjustment module 10 – External display and adjust
Printing date:VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermany39442-EN-130321All statements concerning scope of delivery, application, practi
73 Product descriptionVEGABAR 55 • Probus PA - climate compensated39442-EN-1303213 Product description3.1 CongurationThe scope of delivery encompa
83 Product descriptionVEGABAR 55 • Probus PA - climate compensated39442-EN-130321213Fig. 1: Example of a VEGABAR 55 with process tting G1½ A and pla
93 Product descriptionVEGABAR 55 • Probus PA - climate compensated39442-EN-130321This operating instructions manual applies to the following instrume
Kommentare zu diesen Handbüchern