Operating InstructionsVEGABAR 61Profibus PADocument ID:28251Processpressure/Hydrostaticp

Seite 2 - Contents

of sensor functions (also available as a download).A CD with theappropriate files can be ordered via e-mail under [email protected] orby phone from one

Seite 3

Up to the time of installation, thepackages must be left closed andstored according to the orientation and storage markings on theoutside.Unless other

Seite 4 - 1 About this document

4 Mounting4.1 General instructionsMake sure that all parts of the instrument in contact with the measuredproduct, especially the sensor element, proce

Seite 5 - 2 For your safety

Ventilation for the measuring cell is realised by means of a filterelementin the socket of the electronics housing. Ventilation for theelectronics hous

Seite 6

12Fig. 4: Temperature ranges1 Process temperature2 AmbienttemperatureDanger:Instrumentsin the version "Oil and grease free for oxygen" shoul

Seite 7

4.3 Mounting steps, tube isolating diaphragmThe isolating systems are temperature and pressure-aged at 80 °Cand 18 bar. The zero point is adjusted at

Seite 8 - 3 Product description

90 mm (3.54")R 3,5 mm(0.14")3mm(0.12")70 mm (2.76")8 mm(0.32")93 mm (3.66")110 mm (4.33")Fig. 5: Drilling template

Seite 9 - 3.2 Principle of operation

5 Connecting to power supply5.1 Preparing the connectionAlways keep in mind the following safety instructions:l Connect only in the complete absence o

Seite 10 - 3.3 Operation

In systems with potential equalisation, connectthe cable screendirectly to ground potential at the power supply unit, in the connectionbox and at the

Seite 11 - 28251-EN-081209

11 Tighten thecompression nut of the cable entry. The seal ring mustcompletely encircle the cable12 Screw the housing cover onThe electrical connectio

Seite 12 - 4 Mounting

Contents1 About thisdocument1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41.2 Target group . . . . . . . . . . . . .

Seite 13

2 Remove the housing socket from the mounting plate321Fig. 7: Components of theexternal housing for plics®devices1 Screws2 Wall mounting plate3 Cable

Seite 14 - 4.2 Mounting steps

5.3 Wiring plan, singlechamber housingThe following illustrations apply to the non-Ex as well as to the Ex-iaversion.1 243Fig. 8: Material versions, s

Seite 15 - 4.3 Mounting st

I²CDisplay1 2 5 6 7 83412Fig. 9: Electronics and connectioncompartment, single chamber housing1 Plug connector for VEGACONNECT (I²C interface)2 Spring

Seite 16 - (0.14")

123465Fig. 11: Double chamber housing1 Housingcover, connection compartment2 Blind stopper or plug M12 x 1 for VEGADIS 61 (optional)3 Housing cover, e

Seite 17 - 5 Connecting to power supply

312Display1 2I²CFig. 13: Connection compartment, double chamberhousing1 Plug connector for VEGACONNECT (I²C interface)2 Ground terminal for connection

Seite 18 - 5.2 Connection procedure

5.5 Wiring plan double chamber housing Ex d1 2 345Fig. 15: Double chamber housing1 Housingcover, connection compartment2 Blind stopper or plug M12 x 1

Seite 19

121 2Fig. 17: Connection compartment doublechamber housing Ex d1 Spring-loaded terminals for power supply and cable screen2 Ground terminal for connec

Seite 20

+-12Fig. 19: Wire assignment, connection cable1 brown (+) andblue (-) to power supply or to the processing system2 Shielding5.7 Wiring plan, exte rnal

Seite 21 - 5.3 Wiring plan, single

5 6 7 81 2DisplayI²C45123Fig. 21: Electronics and connectioncompartment1 Plug connector for VEGACONNECT (I²C interface)2 Spring-loaded terminals for c

Seite 22

1 2 3 4 6 3 4 1 2 5Fig. 22: Connection of thesensor in the housing socket1 Brown2 Blue3 Yellow4 White5 Shielding6 Breather capillariesI2CDisplay11

Seite 23

7 Setup with PACTwareand other adjustment programs7.1 Connect the PC via VEGACONNECT . . . . . . . . . . . .467.2 Parameter adjustment with PACTware .

Seite 24

5.8 Switch on phaseAfter VEGABAR 61 is connected to voltage supply or after voltagerecurrence, the instrument carries out a self-check for approx. 30s

Seite 25

6 Set up with the indicating and adjustmentmodule PLICSCOM6.1 Short descriptionThe indicating and adjustment module is used for measured valuedisplay,

Seite 26

Fig. 24: Mounting the indicatingand adjustment moduleNote:If you intend to retrofit the instrument with an indicating and adjustmentmodule for continuo

Seite 27

6.3 Adjustment system1.1231Fig. 25: Indicating and adjustmentelements1 LC display2 Indication of the menu item number3 Adjustment keysl [OK] key:- Mov

Seite 28 - 5 6 7 81 2

6.4 Setup procedureBefore starting the actual parameter adjustment of a Profibus PAsensor, the address setting must first be carried out. You will find a

Seite 29 - 1 2 3 4

1 Push the [OK] button in the measured value display, the menuoverview is displayed.▶Basic adjustmentDisplayDiagnosticsServiceInfo2 Confirm the menu &q

Seite 30 - 5.8 Switch on pha

6 Enter the requested density value with [->] and [+], confirm with[OK] and move to position correction with [->].The adjustment unit is hence sw

Seite 31 - 6.1 Short description

Information:For anadjustment with filling, you simply enter the displayed actualmeasured value. If the adjustment ranges are exceeded, the message"

Seite 32

In the menu items "zero" and "span" you determine the span of thesensor, the span corresponds to the end value.Information:The ste

Seite 33 - - Select editing position

▶Basic adjustmentDisplayDiagnosticsServiceInfo2 Confirm the menu "Basic adjustment" with [OK], the menu item"Unit" will be displaye

Seite 34 - 6.4 Setup procedur

1 About this document1.1 FunctionThis operating instructions manual provides all the information youneed for mounting, connection and setup as well as

Seite 35

Proceed as follows:1 Edit thembar value in the menu item "zero" with [OK].Zero adjustment000.0 %P=+0000.0 mbar0000.0 mbar2 Set the requested

Seite 36

A linearization is necessary for all vessels in which the vessel volumedoes notincrease linearly with the level - e. g. with a cylindrical orspherical

Seite 37 - 0000.0 mbar

Copy sensor data?Basic adjustmentIf the "Reset" is carried out, the sens or resets the values of thefollowing menu items to the reset values

Seite 38

6.5 Menu schematicInformation:Depending on the version and application, the highlighted menuwindows are not always available.Basic adjustment1▶Basic a

Seite 39

Service4Basic adjustmentDisplayDiagnostics▶ServiceInfo4.1Additional PA valueSecondary Value 14.2Out-Scale-UnitVolumel4.3PA-Out-Scale100.00 lin %= 0.0

Seite 40

6.6 Saving the parameter adjustment dataIt is recommended noting the adjusted data, e.g. in this operatinginstructions manual and archive them afterwa

Seite 41 - Linearisation curve

7 Setup with PACTware andother adjustmentprograms7.1 Connect the PC via VEGACONNECT312Fig. 26: Connection of the PC via VEGACONNECT directlyto the sen

Seite 42 - Copy sensor data?

Necessary components:l VEGABAR 61l PC with PACTware and suitable VEGA DTMl VEGACONNECTl Power supply unit or processing system7.2 Parameter adjustment

Seite 43 - 6.5 Menu schem

8 Maintenance and fault rectification8.1 MaintainWhenused in the correct way, no special maintenance is required innormal operation.In some application

Seite 44

? Wrong presentation of the measured value in Simatic S5l Simatic S5 cannot interpret the number format IEEE of themeasured valueà Insert converting c

Seite 45 - 6.6 Saving the pa

2 For your safety2.1 Authorised personnelAll operations described in this operating instructions manual must becarried out only by trained specialist

Seite 46

? E041l Hardwareerror, electronics defectiveà Exchange instrument or return instrument for repair? E113l Communication conflictà Exchange instrument or

Seite 47

Push the button " Update software" and select the previouslyextracted hex file. Then the software update can be started. Furtherfiles are inst

Seite 48 - 8.2 Remove interferences

9 Dismounting9.1 Dismounting stepsWarning:Before dismounting, be aware of dangerous process conditions suchas e.g. pressure in the vessel, high temper

Seite 49

10 Supplement10.1 TechnicaldataGeneral dataPressure type Gauge pressure or gauge pressureMeasuring principle Depending on the measuring range ceramic-

Seite 50 - 8.4 Software update

Dynamic behaviour outputRun-up timeapprox. 10 s90 %100 %10 %ttTAttS21Fig. 28: Sudden change ofthe process variable, dead time tT, rise time tAand step

Seite 51 - 8.5 Instrument repair

Adjustment range of the zero/span adjustment relating to the nominal measuring range:- zero -20 … +95 %- Span -120 … +120 %11)- Difference between zero

Seite 52 - 9 Dismounting

Nominal range Overload, max. pressure Overload, min. pressure0 … 100 bar/0 … 10 MPa 200 bar/20000 kPa0 bar abs.0 … 400 bar/0 … 40 MPa 800 bar/80000 kP

Seite 53 - 10 Supplement

Long-term stability (similar to DIN 16086, DINV 19259-1 and IEC 60770-1)Applies to digital interfaces (HART, Profibus PA, Foundation Fieldbus) as well

Seite 54

Process conditionsThespecifications to the pressure stage and the product temperature are used as an overview.The specifications of the type label are a

Seite 55

Electromechanical data - version IP 66/IP 67Cableentry/plug17)- Single chamber housing l 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm),1 x blind stopp

Seite 56

2.5 Safety approval markings and safety tipsThe safety approval markings and safety tips on the device must beobserved.2.6 CE conformityThis device fu

Seite 57

Connection cable- Configuration fourwires, one suspension cable, one breathercapillary, screen braiding, metal foil, mantle- Wire cross-section 0.5 mm²

Seite 58

Adjustment elements 4 keysProtectionclass- unassembled IP 20- mounted into the sensor without cover IP 40Materials- Housing ABS- Inspection window Pol

Seite 59

ApprovalsDependingon the version, instruments with approvals can have d ifferent technical data.For these instruments, the corresponding approval docum

Seite 60

10.2 Information on Profibus PAInstrumentmaster fileThe instrument master file (GSD) contains t he characteristic data of the Profibus PA instrument.These

Seite 61

bar% Lin%Min-MaxadjustmentLinearizationPROFIBUS PA-outputPrimaryValueSecondaryValue 1SecondaryValue 2TargetModeFailuremodeAlarmsScalingSource forscali

Seite 62

- PA-OUT value ofthe FB1 after scalingl Temperature- PA-OUT value ofthe FB2 after scalingl Additional Cyclic Value- Additional cyclical value (dependi

Seite 63

Byte-No.1 2 3 4 5 6 7 8 9 10FormatValueIEEE-754-FlieskommazahlPA-OUT(FB1)StatusStatus(FB1)IEEE-754-FlieskommazahlTemperature(FB2)StatusStatus(FB2)Fig.

Seite 64

Coding of the status byte associated with the PA output valueStatuscodeDescription according to Pro-fibus standardPossible cause0x00 bad - non-specific

Seite 65 - - PA-OUT value of

10.3 DimensionsHousingin protection class IP 66/IP 67~ 69 mm(2.72")ø 77 mm(3.03")112 mm (4 13/32")M20x1,5/½ NPT~ 69 mm(2.72")ø 77

Seite 66

Housing in protection IP 66/IP 68 (1 bar)117 mm (4.61")120 mm (4.72")~ 103 mm(4.06")~ 105 mm (4.13")ø 77 mm(3.03")116 mm (4.5

Seite 67

2.10 Environmental instructionsProtectionof the environment is one of our most important duties. Thatis why we have introduced an environment manageme

Seite 68 - 10.3 Dimensions

IP 68 version with external housing1265 mm(2 9/16")68 mm(2 43/64")92 mm(3 5/8")42mm(1 21/32")40mm(1 37/64")90 mm (3 35/64&quo

Seite 69

VEGABAR 61 - flangeversion122mm (4 51/64") 122mm (4 51/64") 51mm (2 1/64") 3 5/8"7 1/2" 1/8" 1/8"4xø 5/8"4xø 5/

Seite 70

VEGABAR 61 - flangeversion3 5/8"7 1/2" 1/8" 1/8"4xø 5/8"4xø 5/8"4 3/4"6 63/4"3 7/8"2 9/32" 2 43/64&qu

Seite 71

VEGABAR 61 - tube isolating diaphragm 14 31/64" 1 13/16" 1 1/32"5 33/64" 2 3/64" 1 17/64"6 9/64" 2 3/4" 1 31/3

Seite 72

VEGABAR 61 - tube isolating diaphragm 24 31/64" 5 3/64" 1 1/32"6 9/64" 6 11/16" 1 31/32"6 17/32" 7 11/64" 3 3/

Seite 73

VEGABAR 61 - tube isolating diaphragm 3128 mm (5 3/64")M55x2ø 56mm (2 13/64")ø 44,5mm (3 3/4")~ 132 mm (5 13/64")ø 38,5 mm (1 33/6

Seite 74

VEGABAR 61 - hygienic fittingRV115 mm (4 17/32")ø 78 mm (3 5/64")Fig. 44: VEGABAR 61 - RV = bolting according to DIN 1185176 VEGABAR 61 • Pro

Seite 75

VEGABAR 61 - cell isolating diaphragm135 mm (5 5/16")100 mm (3 15/16")b4 17/32"45/64" 2 43/64"5 29/32"45/64" 3 15/3

Seite 76

VEGABAR 61 - threaded versionSW32SW41SW55SW55G¾BG1B1½”NPTG1½B ø 22 mm (0.87") ø 32 mm (1.26")20 mm(0.79")~105 mm(4.13")~105 mm(4.7

Seite 77

VEGABAR 61 - Threaded versionwith temperature adapterSW27G1Aø 26 mm (1 1/32")LFig. 47: VEGABAR 61 - threaded version withtemperature adapter. The

Seite 78 - GD/GF GN

3 Product description3.1 ConfigurationThe scope of delivery encompasses:l VEGABAR 61 pressure transmitterl Documentation- this operating instructions m

Seite 79

10.4 Industrial property rightsVEGA product lines are global protected by industrial property rights.Further information see http://www.vega.com.Only

Seite 80 - 10.5 Trademark

VEGABAR 61 • Profibus PA 8110 Supplement28251-EN-081209

Seite 81

82 VEGABAR 61 • Profibus PA10 Supplement28251-EN-081209

Seite 82

VEGABAR 61 • Profibus PA 8310 Supplement28251-EN-081209

Seite 83

VEGA Grieshaber KGAm Hohenstein 11377761 SchiltachGermanyPhone +49 7836 50-0Fax +49 7836 50-201E-mail: [email protected] date:ISO 9

Seite 84 - ISO 9001

l Article number instrumentl Technical data: Measuring range, process pressure, processtemperature, signal output, voltage supply, protection, protect