25
7Maintenanceandfaultrectication
VEGACAP 65 • - two-wire
30019-EN-130916
2
1
Fig. 17: Compensation key
1 Measuring range selection switch (compensation key)
2 Control lamp
16. Keep the measuring range selection switch (1) pushed until the
controllamp(2)ashesgreen.
17. Carryouttheadjustmentagain.Seechapter"Set-up, adjustment
elements".
18. Screw the housing cover back on
Theelectronicsexchangeisnownished.
7.4 Shortening of the probe
The probe can be shortened by any amount.
1. Loosen the two pins on the gravity weight (hexagon) and remove
the pins.
2. Pull the cable out of the gravity weight.
3. To avoid splicing of the steel cable, tin the cable before shortening
with a soldering iron and tighten the wire.
4. Shortenthecablewithacut-owheelormetalsawatthelower
end. Make sure the length is correct before shortening.
70 mm
(2.76")
200 mm (7.87")
Fig. 18: Take the gravity weight into account and shorten the cable respectively
5. Shiftthegravityweightagainushtothecableandxitwiththe
two pins.