Vega D81 Hydrostatic pressure transmitter Bedienungsanleitung Seite 30

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 44
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 29
30 Pressure transmitter D81
Setup
HART
®
menu schematic (continuation)
1.3.3
1.3.4
6 Write protect None
7 Descriptor
8 Message
9 PV Snsr s/n
Final asmbly num 0
Revision #’s
1.4
like menu 1.3.4
1.2.3.2
like display 4.1
like display 4.2
1.2.3
1.2.3.1
Safety
enquiry
acknowl-
edgement
Indication of the
sensor measuring
range limits
Indication of the
sensor measuring
range limits
(1.3.4.5) Date adjustment
individual text adjustment
Measured value
Modification of the unit
Info on lower and upper measur-
ing range limit as well as min.
span (also the sensor characteris-
tics values)
1.4.1
1.4.2
Adjustment of the integration time
Adjustment of zero and span
Adjustment of zero and span
linear
Indication of the level in % of the
adjusted span
1.4.3
Menu fields (broken line):
repetition of previous page
}
1.3.3.1
1.3.3.2
Indication of the sensor serial number (as under menu window 1.3.4.3)
Unknown Enumerator
General
individual text adjustment
None
Empty adjustment with
medium
Full adjustment with
medium
(1.3.4.4) Measurement loop designation as under menu window 1.3.1
(1.3.4.3) Indication of the sensor serial number
}
1.2.3.1
Generic: D84
Range values
1 PV LRV 0.000mbar
2 PV URV 95.981mbar
3 PV LSL 0.000mbar
4 USL 100.000mbar
HELP HOME
Generic: D84
Device information
1 Distributor
2 Model Generic
3 Dev id 11017815
4Tag D84
5 Date 01/01/97
HELP SAVE HOME
Generic: D84
Detailed setup
1 Sensors
2 Signal condition
3 Output condition
4 Device information
SAVE HOME
Generic: D84
PV LRV
0.000 mbar
0.000
HELP DEL ESC ENTER
Generic: D84
PV URV
100.000 mbar
100.000 mbar
HELP DEL ESC ENTER
Generic: D84
Sensors
1 PV 97.214 mbar
2 PV Snsr unit mbar
3 Sensor information
HELP SAVE HOME
Generic: D84
Signal condition
1 Snsr Damp 1.000 s
2 URV 100.000 mbar
3 AI LRV 0.000 mbar
4 Xfer fnctn Linear
5 AI % rnge 97.101 %
HELP SAVE HOME
Generic: D84
Output condition
1 Analog output
2 HART output
SAVE HOME
Generic: D84
Enter values
1 PV LRV 0.000 mbar
2 URV 100.000 mbar
3 PV USL
4 PV LSL 0.000 mbar
HELP HOME
Generic: D84
Calibration
1 Apply values
2 Enter values
HELP SAVE HOME
Generic: D84
Apply new 20 mA input
ABBR. OK
Generic: D84
Apply new 4 mA input
ABORT OK
Generic: D84
Set the:
14 mA
2 20 mA
3Exit
ABORTENTER