12 Pressure transmitter D85
Protective measures
1)
Protection IP 66, IP 67, IP 68
VEGABOX 01 IP 66 and IP 67
Protection class III
Overvoltage category III
Accuracy (similar to DIN 16 086, DIN V 19 259 - 1 and IEC 770)
Deviation
Reference conditions (acc. to IEC 770)
- temperature 15°C 30°C
- humidity 45 % 75 %
- air pressure 86 kPa 106 kPa
Determination of characteristics limit point adjustment acc. to DIN 16 086
Characteristics linear
Deviation in characteristics
- Turn Down 1 : 1 < 0.25 % with accuracy class 0.25
< 0.1 % with accuracy class 0.1
- Turn Down up to 1 : 5 typ. < 0.3 % with accuracy class 0.25
typ. < 0.1 % with accuracy class 0.1
- Turn Down up to 1 : 10 typ. < 0.4 % with accuracy class 0.25
typ. < 0.2 % with accuracy class 0.1
Hysteresis < 0.05 %
Repeatability < 0.05 %
Influence of the ambient temperature
Average temperature coefficient
of the zero signal
2) 3)
- Turn Down 1 : 1 < 0.15 %/10 K with accuracy class 0.25
< 0.05 %/10 K with accuracy class 0.1
- Turn Down up to 1 : 5 typ. < 0.225 %/10 K with accuracy class 0.25
typ. < 0.075 %/10 K with accuracy class 0.1
- Turn Down up to 1 : 10 typ. < 0.3 %/10 K with accuracy class 0.25
typ. < 0.1 %/10 K with accuracy class 0.1
Long-term stability
Long-term stability of the zero signal
4)
< 0.25 % per 2 years
Other actuating variables
Calibration position upright, diaphragm points downwards
Influence of the installation position < 5 mbar
Vibration resistance mechanical vibrations with 4 g and
5 100 Hz, tested acc. to the regulations of
German Lloyd GL-characteristics 2
Product description
1)
Maintaining the housing protection IP 66 or IP 67 requires the use of a seal that correctly fits the cable in
the cable entry. If the supplied seal does not fit, a suitable seal must be provided by the customer.
2)
Relating to the nominal range.
3)
In compensated temperature range of 0°C +80°C, reference temperature 20°C.
4)
Acc. to IEC 770, item 6.3.2 relating to the nominal range.
Seitenansicht 11