Pressure transmitter D96/D97 (Profibus PA) 21
Setup
ti
1
s
Adjustment without taking the current
pressure into account (dry adjustment)
Adjustment without pressure comprises four
steps:
1 Selection of the unit in which the adjustment
is to be carried out
2 Offset correction
3 Adjustment of zero
4 Adjustment of span
The offset correction (only in gauge pressure
mode) defines the reference positions for the
measurement. It can be carried out:
- before or after adjustment of zero and
span
- before or after installation of transmitters.
The transmitter must be unpressurised for
the offset correction!
Adjustment without pressure can be carried
out with the instrument mounted or dis-
mounted (e.g. workroom). Any ambient pres-
sure has no effect on the adjustment.
1 Selection of the unit
Choose the unit with the "+" or ""
key. Save the selected unit with
the "OK" key.
2 Adjustment of zero
Adjust zero with the "+" or "" key
and save with the "OK" key.
Zero
0,0
mbar
Span
100,0
mbar
4 Adjustment of span
Adjust span with the "+" or "" key
and save with the "OK" key.
Note:
- A modification of zero does not influence
the span.
- By pushing the "+" and "" keys, the output
remains on the last value - only after saving
with the "OK" key will the output take on the
adjusted value.
Evaluation
Linearisation curve
Is only displayed if a linearisation
curve has been activated with VVO.
Adjustment of the integration time
An integration time t
i
of 0 s 50 s
can be adjusted with the "+" or ""
key for damping of pressure
shocks. The adjusted value can be saved
with the "OK" key.
Selection of the indicated unit
The measurement unit for the indi-
cated measured value as well as
for the display module can be
selected here (with the "+" or ""
key and the "OK" key). If you select
"scal.", the scaled PA output value
will be displayed (as adjusted
under "Scaling outp.").
Lin.-
curve
activ
Adjust
without
pres.
Unit
0,0
mbar
psi
kPa
inHg
Unit
0,0
skal
psi
kPa
%
m
ft
inch
mbar