Cyclicaldatatrac
The master class 1 (e.g. PLC) cyclically reads out measured values from the sensor during opera-
tion. The below block diagram below shows which data can be accessed by the PLC.
1
Electronic
Te mperature
Count
Rate
Primary
Value
Fig. 19: POINTRAC 31: Block diagram with AI FB 1 … AI FB 3 OUT values
TB Transducer Block
FB 1 … FB 3 Function Block
1 Output value, adjustable (depending on instrument type and application - see following list)
•
Process temperature
•
Vapour density
•
Density
•
Volume ow
•
Ground
•
Belt speed
•
Loading
Module of the PA sensors
For the cyclic data trac, POINTRAC 31 provides the following modules:
•