4 Setup
Electrical connection and setup can be sim-
plified by temporarily mounting the cover or
the display module on VEGABAR laterally or
downwardly displaced.
OPERATE
t
SPAN
ZERO
i
Key +: Modify
value (increase
Key –: Modify
value (de-
crease)
Setup
Display module
VEGABAR
87654321
4 ¼ 20 mA
DISPLAY
VEGADIS 10
12 ¼ 36 V DC E12
+ -
+
-
OK
Tighten screws lightly
4.1 Display module
Digital value
- 4 digital and decimal point
- basic function module: indicates a perma-
nent bar measurement range
- menu-driven module: allows individual
scaling of bar measurement range
Connection is made via a plug connector
according to the "Electrical connection“ sec-
tion.
4.2 Setup with module "Adjustment
of the basic functions“
Adjustment elements
Tendency
indication
Bar graph
Digital value
Adjustment system
Select the required function with the rotary
switch.
Change the value with the "+“ and "–“ keys.
Reset rotary switch to OPERATE, the ad-
justed values are transferred to the EEP-
ROM-memory. They remain there even in
case of voltage drop.
Adjustment
For adjustment of zero and span, an amme-
ter must be connected to terminals 1 and 3.
The measured value is identical to the output
current.
1 Adjustment of zero
(e.g. process pressure zero or vessel
empty)
Set rotary switch to zero
By pushing the "+“ and "–“ keys (simulta-
neously) the current jumps directly to 4 mA
or set it to 4 mA by pushing the "+“ and "–“
keys.
Adjustment range of zero:
-20 % … +95 % of the nominal range (corre-
sponds to turn of up to +95 %)
Rotary switch: Select requested
function
Seitenansicht 16