Vega VEGABAR 44 HART Bedienungsanleitung Seite 34

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 44
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 33
34 VEGABAR 44 (HART
®
)
Setup
HART
®
menu schematic (continuation)
1.3.3
1.3.4
1.4
like menu 1.3.4
1.2.3.2
like display 4.1
like display 4.2
1.2.3
1.2.3.1
Generic: SENSOR
Set the:
14mA
2 20mA
3Exit
ABORT ENTER
Safety
enquiry
Acknow-
ledge
Indication of the
sensor measuring
range limits
Generic: SENSOR
PV LRV
0.000 mbar
10.000 mbar
HELP DEL ESC ENTER
Generic: SENSOR
PV URV
100.000 mbar
90.300 mbar
HELP DEL ESC ENTER
Indication of the
sensor measuring
range limits
(1.3.4.5) Date adjustment
individual text adjustment
Generic: SENSOR
Sensors
1 PV 80.945 mbar
2 PV Snsr unit bar
3 Sensor information
HELP SAVE HOME
Measured value
Modification of the unit
Info on lower and upper
measuring range limit as well
as min. span (also the sensor
characteristics values)
1.4.1
Generic: SENSOR
Signal condition
1 Snsr Damp 1.000 s
2 AI URV
3AI LRV
4 Xfer fnctn
5 % rnge
HELP SAVE HOME
1.4.2
Adjustment of the integration time
Adjustment of zero and span
Adjustment of zero and span
linear
Indication of the level in % of the
adjusted span
Generic: SENSOR
Output condition
1 Analog output
2 HART output
SAVE HOME
1.4.3
Menu fields (broken line:
Repetition of previous page
}
1.3.3.1
1.3.3.2
Indication of the sensor serial number (as menu window 1.3.4.3)
Unknown Enumerator
General
Generic: SENSOR
Enter values
1 PV LRV bar
2 PV URV bar
3 PV USL bar
4 PV ´LSL bar
HELP HOME
Generic: SENSOR
Apply new 4 ma input
ABORT OK
individual text adjustment
None
Empty adjustment with
medium
Full adjustment with
medium
(1.3.4.4) Measurement loop designation as under menu window 1.3.1
(1.3.4.3) Indication of the sensor serial number
}
A1
B1
1.2.3.1
Generic: SENSOR
Calibration
1 Apply values
2 Enter values
HELP SAVE HOME
Generic: SENSOR
Apply new 20 ma input
ABORT OK
Generic: SENSOR
Range values
1 PV LRV bar
2 PV URV bar
3 PV LSL bar
4 PV USL bar
HELP HOME
Generic: SENSOR
Device information
1 Distributor
2 Model
3Dev id
4Tag
5 Date 10/01/97
HELP SAVE HOME
6 Write protect None
7 Descriptor
8 Message
9 PV Snsr s/n
Final asmbly num 0
Revision #´s
Generic: SENSOR
Detailed setup
1 Sensors
2 Signal condition
3 Output condition
4 Device information
SAVE HOME
Seitenansicht 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 43 44