Ventilation for the measuring cell is realised by means of a lter
element
in the socket of the electronics housing. Ventilation for the
electronics housing is realised via an additional lter element in the
area of the cable glands.
2
1
2
2
2
1
1
1
Fig. 3: Position of the lter
elements
1 Filter element for ventilation of the electronics housing
2 Filter element for ventilation of the measuring cell
Informa
tion:
Mak
e
sure that the lter elements are always free of buildup during
operation. A pressure washer must not be used for cleaning.
With instrument versions in protection IP 66/IP 68, 1 bar, the ventilation
is realised via the capillaries in the x connected cable. The lter
elements are replaced by blind stoppers.
Higher process temperatures often mean also higher ambient
temperatures. Make sure that the upper temperature limits stated in
chapter "Technical data" for the environment of the electronics housing
and connection cable are not exceeded.
Ventilation
Temperature limits
12 VEGABAR 65 4 20 mA/HAR