29
5 Connecting to power supply
VEGABAR 66 • 4 ... 20 mA/HART
36737-EN-130321
1234
6
3
4
1
2
5
Fig. 25: Connection of the sensor in the housing socket
1 Brown
2 Blue
3 Yellow
4 White
5 Shielding
6 Breather capillaries
I
2
C
Display
1
12 5 678
Fig. 26: Wiring plan external electronics
1 Voltage supply
5.8 Switch-on phase
After connecting VEGABAR 66 to power supply or after a voltage
recurrence, the instrument carries out a self-check for approx. 30
seconds:
•
Internal check of the electronics
•
Indicationoftheinstrumenttype,thermwareaswellasthesen-
sor TAGs (sensor designation)
•
Outputsignaljumpsbriey(approx.10seconds)tothesetfault
current
Terminal compartment,
housing socket
Wiring plan e
xternal
electronics
Switch-on phase