Vega VEGABAR 66 Profibus PA - climate compensated Bedienungsanleitung Seite 54

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 76
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 53
54
10 Supplement
VEGABAR 66 • Probus PA - climate compensated
36792-EN-130321
Ʋ Ohmic contact Between ground terminal, process tting and transmitter
Ʋ Connection cable between IP 68
housing and external electronics
PUR, FEP, PE
Ʋ Type label support with IP 68 version
on cable
PE hard
Weight approx.
Ʋ Basic weight 0.7 kg (1.543 lbs)
Ʋ Suspension cable 0.1 kg/m (0.07 lbs/ft)
Ʋ Connection tube 1.5 kg/m (1 lbs/ft)
Ʋ Straining clamp 0.2 kg (0.441 lbs)
Ʋ Threaded tting 0.4 kg (0.882 lbs)
Lengths
Ʋ Connection tube 0.25 … 6 m (0.82 … 19.69 ft)
Output variable
Output signal digital output signal, format according to IEEE-754
Sensor address 126 (default setting)
Current value
10 mA, ±0.5 mA
Dynamic behaviour output
Run-up time approx.
10 s
90 %
100 %
10 %
t
t
T
A
t
t
S
2
1
Fig. 31: Sudden change of the process variable. t
T
: dead time; t
A
: rise time; t
S
: jump response time
1 Process variable
2 Output signal
Dead time ≤ 150 ms
Rise time ≤ 100 ms (10 … 90 %)
Step response time ≤ 250 ms (ti: 0 s, 10 … 90 %)
Damping (63 % of the input variable) 0 … 999 s, adjustable
Additional output parameter - temperature
Processing is made via output signal HART multidrop, Probus PA and Foundation Fieldbus
Range
-50 … +150 °C (-58 … +302 °F)