6 Set up
6.1 Setup steps w
ithout VEGADIS 12
After mounting and electrical connection, VEGABAR 75 is
ready for operation.
à Switch on voltage
The electronics now carries out a self-check for approx. 2
seconds. Then VEGABAR 75 delivers a current of 4 20 mA
according to the actual level.
6.2 Setup steps with VEGADIS 12
l zero - measuring range begin
l span - meas
uring range end
l ti - Integration
time
VEGADIS 12
TRANSMITTER
+
4...20mA
+
not used
8765
1211101 2 3
-
-
ti
ZERO
SPAN
TRANSMITTER
1
2
3
Fig. 8: Adjustment elements of VEGADIS 12
1 Rotary switch: choose the requested function
2 [+] key, change value (rising)
3 [-] key, change value (falling)
l With
the rotary switch the requested function is selected
l With the [+] and [-] keys, the signal current
or the
integration time are adjusted
l Finally the rotary switch is set to position "OPERATE"
The set values are transmitted to the EEPROM memory and
remain there even in case of voltage loss.
Proceed as follows for adjustment with VEGADIS 12:
1 Open housing cover
2 Connect hand multimeter to terminals 10 and 12
3 Meas. range begin: Set rotary switch to "zero"
Adjustment volume
Adjustment system
Adjustment steps, adjustment
VEGABAR 75 - 4 20 mA/HART 19
Set up
28433-EN-070718
Seitenansicht 18
1 2 ... 14 15 16 17 18 19 20 21 22 23 24 ... 39 40

Kommentare zu diesen Handbüchern

Keine Kommentare