Vega VEGAFLEX 86 4 … 20 mA_HART - four-wire Bedienungsanleitung Seite 37

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 92
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 36
37
6 Set up with the display and adjustment module
VEGAFLEX 86 • 4 … 20 mA/HART - four-wire
41845-EN-130314
Enter the appropriate distance value in m (corresponding to the
percentage value) for the full vessel. The distance refers to the sensor
reference plane (seal surface of the process tting). Keep in mind that
the max. level must lie below the dead band.
In this menu item you can enter the min. adjustment for the level. With
interface measurement this is the minimum total level.
Adjust the requested percentage value with [+] and save with [OK].
Enter the suitable distance value in m for the empty vessel (e.g.
distance from the ange to the probe end) corresponding to the per-
centage value. The distance refers tot he sensor reference plane (seal
surface of the process tting).
This menu item is only available if you have selected interface meas-
urement under the menu item "Application".
You can accept the adjustment of the level measurement also for the
interface measurement. If you select "Yes", the current setting will be
displayed.
If you have selected "No", you can enter the adjustment for the inter-
face separately. Enter the requested percentage value.
Setup - Min. adjustment
Leve
l
Setup - Max. adjustment -
Interface
Seitenansicht 36
1 2 ... 32 33 34 35 36 37 38 39 40 41 42 ... 91 92

Kommentare zu diesen Handbüchern

Keine Kommentare