49
11 Supplement
POINTRAC 31 • 8/16 mA/HART - four-wire
43388-EN-131119
Relay output
Output Relayoutput(SPDT),oatingspdt
Switching voltage
Ʋ Min. 10mV
Ʋ Max. 253VAC,253VDC
Switching current
Ʋ Min. 10 µA
Ʋ Max. 3 A AC, 1 A DC
Breaking capacity
Ʋ Min. 50 mW
Ʋ Max. 750VAAC,40WDC
If inductive loads or stronger currents are switched
through, the gold plating on the relay contact surface will
be permanently damaged. The contact is then no longer
suitable for switching low-level signal circuits.
Contact material (relay contacts) AgNi or AgSnO and Au plated
Accuracy (according to DIN EN 60770-1)
Process reference conditions according to DIN EN 61298-1
Ʋ Temperature +18…+30°C(+64…+86°F)
Ʋ Relative humidity 45 … 75 %
Ʋ Air pressure 860 … 1060 mbar/86 … 106 kPa (12.5 … 15.4 psig)
Repeatability ≤0.5%
Deviation with bulk solids The values depend to a great extent on the application.
Bindingspecicationsarethusnotpossible.
DeviationunderEMCinuence ≤1%
Variablesinuencingmeasurementaccuracy
Specicationsapplyalsotothecurrentoutput
Temperature drift - Current output ±0.03 %/10 K relating to the 16 mA span max. ±0.3 %
Deviation on the current output through
analogue/digital conversion
<±15 µA
Deviation on the current output due to
strong, high frequency electromagnetic
interference acc. to EN 61326
<±150 µA
Ambient conditions
Ambient, storage and transport tempera-
ture
-40…+60°C(-40…+140°F)
Process conditions
Fortheprocessconditions,pleasealsonotethespecicationsonthetypelabel.Thelowervalue
always applies.
Process pressure Unpressurized