Seitenansicht 62
10.2 Information on Found
ation Fieldbus
Block diagram, measured value processing
The following illustration shows the Transducer Block and Function block in simplied form.
t
i
Simulate
ALARM
S
OutPV
FIELD-VALUE
M
od
e & Stat
us
PV
F-TIME
LOW-CUTOFF
L-TYPE
OUT-
SCALE
PV-SCALE
DIRECT
INDIRECT
INDIRECT
SQRT
bar % Lin%
Primary
Value
Secondary
Value 1
Secondary
Value 2
Sensor mounting
correction
AI 2
AI 3
AI 1
TB
Temperature
Sensor_Value
Linearizationadjustment
Channel = 1: Primary Value
Channel = 2: Secondary Value 1
Channel = 3: Secondary Value 2
Channel = 4: Temperature
Fig. 29: Transducer Block VEGABAR 64
TB Transducer Block
AI Function Block (AI =Analogue Input)
Diagram, adjus
tment
The following illustration shows the function of the adjustment:
VEGABAR 64 Foundation Fieldbu
s 63
10 Supplement
28259-EN-081209