6 Setup - radar sensors from software 4.0.0
PLICSCOM •
36433-EN-130228
In this menu item, the internal clock of the sensor is adjusted.
The instrument settings are copied with this function. The following
functions are available:
Store data from the sensor into the indicating/adjustment module.
Store data from the display and adjustment module in the sensor
The following data or settings for adjustment of the display and ad-
justment module are saved:
All data of the menu "Setup" and "Display"
In the menu "Additional settings" the items "Distance unit, tem-
perature unit and linearization"
Thevaluesoftheuserprogrammablelinearizationcurve
The copied data are permanently saved in an EEPROM memory in
the display and adjustment module and remain there even in case of
power failure. From there, they can be written into one or more sen-
sors or kept as backup for a possible sensor exchange.
The type and the volume of the copied data depend on the respective
sensor.
Note:
Before the data are stored in the sensor, a check is carried out to
determineifthedatatthesensor.Ifthedatadonott,afaultsignal
is triggered or the function is blocked. When data are being written
into the sensor, the display shows which instrument type the data
originate from and which TAG-no. this sensor had.
In this menu, you read out the instrument name and the instrument
serial number:
Additional adjustments/
Date time
Additional adjustments/
Copy instrument settings
Info/Instrument name