9 Diagnostics and service
POINTRAC 31 • 8/16 mA/HART - four-wire
39411-EN-131119
Maintenance:Duetoexternalinuences,theinstrumentfunction
islimited.Themeasurementisaected,butthemeasuredvalueis
still valid. Plan in maintenance for the instrument because a failure is
expected in the near future (e.g. due to buildup).
This status message is inactive by default. It can be activated by the
user via PACTware/DTM or EDD.
The following table shows the error codes and text messages in the
status message "Failure" and provides information on causes as well
as corrective measures.
Example for a failure message
Code
Text mes-
sage
Cause Rectication
F008
Error multi
sensor com-
munication
Additional sensors not
switched on
EMCinuences
No other sensor available
Check wiring between the
sensors
Connect the sensors cor-
rectly and make them ready
for operation
F013
Sensor sig-
nals a fault
Error on the current input
No valid measured value
Connected instruments
without function
Check current input
Check connected instru-
ments (Slaves)
F016
Adjustment
data ex-
changed
Valuesofthemin.andmax.
adjustment exchanged
Correct adjustment data
F017
Adjustment
span too small
The values of the min. and
max. adjustment are too
close together
Correct adjustment data
F025
Invalid lineari-
zationtable
Wrong value in the lineari-
zationtable
Correctlinearizationtable
F030
Process value
out of limits
Process values are not
within the adjusted measur-
ing range
Repeat adjustment
F034
EPROM hard-
ware error
Electronics defective Exchanging the electronics
F035
EPROM data
error
Error in the internal instru-
ment communication
Carry out a reset
Exchanging the electronics