Vega VEGABAR 51 Profibus PA Bedienungsanleitung Seite 50

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 76
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 49
50
10 Supplement
VEGABAR 51 • Probus PA
36714-EN-130321
Deviation determined according to the limit point method according to IEC 60770
10)
Applies to the digital signal output (HART, Probus PA, Foundation Fieldbus) as well as to ana-
logue
current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Deviation
Ʋ Turn down 1 : 1 < 0.2 %
Ʋ Turn down up to 5 : 1 < 0.2 %
Ʋ Turn down up to 10 : 1 < 0.3 %
Inuenceoftheproductorambienttemperature
Thermal change zero signal and output span
Applies to the digital signal output (HART, Probus PA, Foundation Fieldbus) as well as to ana-
logue
current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Thermal change zero signal and output span, reference temperature 20 °C (68 °F):
Ʋ In the compensated temperature
range 0 … +100 °C (+32 … +212 °F)
< 0.05 %/10 K x TD
Ʋ Outside the compensated tempera-
ture range
typ. < 0.05 %/10 K x TD
Thermal change, current output
Applies also to the analogue
4 … 20 mA current output and refers to the set span.
Thermal change, current output < 0.05 %/10 K, max. < 0.15 %, each with -40 … +80 °C
(-40 … +176 °F)
0,15 %
-0,15 %
-40°C -20°C
20°C
40°C 60°C 80°C
Fig. 27: Thermal change, current output
Long-termstability(accordingtoDIN16086andIEC60770-1)
Applies to digital interfaces (HART, Probus PA, Foundation Fieldbus) as well as to analogue cur-
rent output 4 … 20 mA. Specications refer to the set span. Turn down (TD) = nominal measuring
range/set span.
Long-term drift of the zero signal
< (0.1 % x TD)/year
Ambient conditions
Ambient, storage and transport temperature
10)
Incl. non-linearity, hysteresis and non-repeatability.
Seitenansicht 49
1 2 ... 45 46 47 48 49 50 51 52 53 54 55 ... 75 76

Kommentare zu diesen Handbüchern

Keine Kommentare