Nominal
range Overload, max. pressure Overload, min. pressure
-15 25 psig 700 psig -15 psig
-15 70 psig 950 psig -15 psig
-15 150 psig 1300 psig -15 psig
-15 350 psig 1900 psig -15 psig
-15 900 psig 2900 psig -15 psig
-1.5 1.5 psig 500 psig -15 psig
-3 3 psig 500 psig -12 psig
-7 7 psig 700 psig -15 psig
Reference conditions and actuating variables (similar to DIN EN 60770-1)
Reference conditions according to DIN EN 61298-1
- Temperature +15 +25 °C (+59 +77 °F)
- Relative humidity 45 75 %
- Air pressure 860 1060 mbar/86 106 kPa (12.5 15.4 psig)
Determination of characteristics Limit point adjustment according to IEC 61298-2
Characteristics curve linear
Reference installation position upright, diaphragm points downward
Inuence of the installation position < 0.2 mbar/20 Pa (0.003 psig)
Deviation determined according to the limit point method according to IEC 60770
8)
Applies to digital HART interface as well as to analogue current output 4 20 mA. Specications
refer to the set span. Turn down (TD) is the relation nominal measuring range/set span.
Deviation < (0.05 + 0.025 x TD)%
Inuence of the product or ambient temperature
Average temperature coecient of the zero signal
Applies to digital HART interface as well as to analogue current output 4 20 mA. Specications
refer to the set span. Turn down (TD) is the relation nominal measuring range/set span.
Nominal measuring range in
bar/kPa
Nominal
measuring
range in psig
Average temperature coe-
cient of the zero signal
9)
Average temperature coe-
cient of the zero signal
10)
0 0.2 bar/0 20 kPa 0 3 psig < (0.05 + 0.05 x TD)%/10K < (0.05 + 0.1 x TD)%/10K
0 0.4 bar/0 40 kPa 0 6 psig < (0.05 + 0.05 x TD)%/10K < (0.05 + 0.1 x TD)%/10K
0 1 bar/0 100 kPa 0 15 psig < (0.02 + 0.05 x TD)%/10K < (0.02 + 0.1 x TD)%/10K
0 2.5 bar/0 250 kPa 0 35 psig < (0.02 + 0.05 x TD)%/10K < (0.02 + 0.1 x TD)%/10K
0 5 bar/0 500 kPa 0 70 psig < (0.02 + 0.03 x TD)%/10K < (0.04 + 0.06 x TD)%/10K
48 VE
GABAR 52 4 20 mA/HART - climate compensated
10 Suppl
ement
35075-EN-081209