Vega VEGABAR 66 Foundation Fieldbus - climate compensat Bedienungsanleitung Seite 34

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 76
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 33
34
6 Set up with the display and adjustment module PLICSCOM
VEGABAR 66 • Foundation Fieldbus - climate compensated
36793-EN-130321
Information:
The steps 1, 3 and 4 are not necessary for instruments which are
alreadypresetaccordingtocustomerspecications!
Youcanndthedataonthetypelabelontheinstrumentorinthe
menu items of the min./max. adjustment.
The display and adjustment module enables the adjustment without
llingorpressure.Thankstothis,youcancarryoutyoursettings
already in the workshop without the instrument having to be installed.
The actual measured value is also displayed in the menu items for
min./max. adjustment.
In this menu item you select the adjustment unit as well as the unit for
the temperature indication in the display.
To select the adjustment unit (in the example switching over from bar
to mbar), proceed as follows:
4)
1. Push the [OK] button in the measured value display, the menu
overview is displayed.
Basic adjustment
Display
Diagnostics
Service
Info
2. Conrmthemenu"Basic adjustment"with[OK], the menu item
"Unit"willbedisplayed.
Unit
Unitofmeasurement
bar
Temperature unit
°C
3. Activate the selection with [OK]andselect"Units of measure-
ment with [->].
4. Activate the selection with [OK] and select the requested unit with
[->] (in the example mbar).
5. Conrmwith[OK] and move to position correction with [->].
The adjustment unit is thus switched over from bar to mbar.
Information:
When switching over to adjustment in a height unit (in the example
from bar to m), the density also has to be entered.
Proceed as follows:
1. Push the [OK] button in the measured value display, the menu
overview is displayed.
2. Conrmthemenu"Basic adjustment"with[OK], the menu item
"Units of measurement"willbedisplayed.
3. Activate the selection with [OK] and select the requested unit with
[->] (in the example m).
Select unit
4)
Selection options: mbar, bar, psi, Pa, kPa, MPa, inHg, mmHg, inH
2
O,
mmH
2
O.
Seitenansicht 33
1 2 ... 29 30 31 32 33 34 35 36 37 38 39 ... 75 76

Kommentare zu diesen Handbüchern

Keine Kommentare