9 Diagnosis, asset management and service
VEGABAR 81 • Probus PA
45020-EN-131107
Code
Text mes-
sage
Cause Rectication
F265
Measurement
function dis-
turbed
Sensor no longer carries
out a measurement
Carry out a reset
Disconnect operating volt-
age briey
F266
Impermissi-
ble operating
voltage
Operating voltage below
specied range
Check electrical connection
if necessary, increase
operating voltage
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause Rectication
C700
Simulation ac-
tive
A simulation is active Finish simulation
Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "Out of specication" and provides information on
causes as well as corrective measures.
Code
Text mes-
sage
Cause Rectication
S600
Impermissi-
ble electronics
temperature
Temperature of the elec-
tronics in the non-specied
section
Check ambient temperature
Isolate electronics
Use instrument with higher
temperature range
S603
Impermissi-
ble operating
voltage
Operating voltage below
specied range
Check electrical connection
if necessary, increase
operating voltage
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause Rectication
M500
Error in the
delivery sta-
tus
Stored delivery status is
incorrect
Send instrument for repair
Function check
Outofspecication
Maintenance