77
10 Supplement
VEGADIF 65 • 4 … 20 mA
36236-EN-130417
Ʋ External housing – 2xcableglandM20x1.5(cable:ø5…9mm),1x
blindplugM20x1.5
or:
– 1xcableglandM20x1.5,1xplug(dependingonver-
sion),1xblindplugM20x1.5
Spring-loadedterminalsforwirecross-
section up to
2.5 mm² (AWG 14)
Display and adjustment module
Voltage supply and data transmission through the sensor
Indication LC display in dot matrix
Adjustment elements 4 keys
Protection rating
Ʋ unassembled IP 20
Ʋ mountedintothesensorwithoutcover IP 40
Material
Ʋ Housing ABS
Ʋ Inspectionwindow Polyester foil
Voltage supply
Operating voltage
Ʋ Non-Ex instrument 12 … 36 V DC
Ʋ Ex-ia instrument 12 … 30 V DC
Ʋ Exd instrument 18 … 36 V DC
Operatingvoltagewithilluminateddisplayandadjustmentmodule
Ʋ Non-Ex instrument 20 … 36 V DC
Ʋ Ex-ia instrument 20 … 30 V DC
Ʋ Exd instrument 20 … 36 V DC
Permissible residual ripple
Ʋ <100Hz U
ss
< 1 V
Ʋ 100Hz…10kHz U
ss
< 10 mV
Load see diagram