8 Diagnosis, Asset Management and service
VEGAPULS 62 • Modbus and Levelmaster protocol
41362-EN-121011
Code
Text mes-
sage
Cause Rectication PA DevS-
pec
Diagnosis
F260
Error in the
calibration
Error in the calibra-
tion carried out in the
factory
Error in the EEPROM
Exchanging the elec-
tronics
Send instrument for
repair
Bit 9
F261
Error in the
congura-
tion
Error during setup
False signal suppres-
sion faulty
Error when carrying
out a reset
Repeat setup
Repeat reset
Bit 10
F264
Installa-
tion/Setup
error
Adjustment not within
the vessel height/
measuring range
Max. measuring range
of the instrument not
sucient
Check or correct
installation and/or
parameter adjustment
Use an instrument
with bigger measuring
range
Bit 11
F265
Meas-
urement
function
disturbed
Sensor no longer
carries out a measure-
ment
Operating voltage
too low
Check operating
voltage
Carry out a reset
Separate operating
voltage briey
Bit 12
The following table shows the error codes and text messages in the
status message "Function check" and provides information on causes
as well as corrective measures.
Code
Text mes-
sage
Cause Rectication
C700
Simulation ac-
tive
A simulation is active Finish simulation
Wait for the automatic end
after 60 mins.
The following table shows the error codes and text messages in the
status message "Outofspecication" and provides information on
causes as well as corrective measures.
Code
Text mes-
sage
Cause Rectication
S600
Unpermissi-
ble electronics
temperature
Temperature of the elec-
tronics in the non-specied
section
Check ambient temperature
Isolate electronics
Use instrument with higher
temperature range
S601
Overlling
Danger of vessel overlling Make sure that there is no
further lling
Check level in the vessel
Function check
Outofspecication
Seitenansicht 52