11 Anexo
POINTRAC 31 • Foundation Fieldbus
41779-ES-131223
FF descriptor Rel. In-
dex
Description Unit
PMT_VOLTAGE_CALI-
BRATION
43 Photomultiplier voltage on delivery -
CORRELATION_CO-
EFF
44 Correlationcoecientforlinearizertable -
ERROR_TEXT 45 Error text -
PMT_VOLTAGE_AC-
TUAL
46 Current photomultiplier voltage -
STANDARDIZATION_
FACTOR
47 Factor for the real value correction -
SERIALNUMBER 48 Serial number -
NAMUR_STATE 49 NAMUR state -
NULL_COUNT_RATE 50 Zero count rate -
COUNT_RATE_PMT 51 Pulse rate photomultiplier (raw values) -
ADJ_DENSITY_ABS_
COEFF
52 Processabsorptioncoecient -
DEV_SW_VER_ASCII 53 Software version -
POINT_LEVEL_AD-
JUST_MODE
54 Point level adjustment mode -
RELAY_VALUE_SEL 55 Relay basic value -
DIGITAL_IN 56 Frequency of digital input -
DIGITAL_IN_BOOL 57 State of digital input -
ANALOG_IN 58 Current on analog input -
FF_CHANNEL_AVAI-
LABLE
59 Available channels -
FF_CHANNEL_USED 60 Used channels -
Mapping of Process Value Status
Hex Quality Sub-Status Condition
0x00 Bad Non-specic Unexpected error
0x01 Bad CongurationError It was tried to set a wrong unit with FF interface
On user error codes: 16, 17, 25, 52, 57, 66, 72,
117, 120
0x0C Bad Device Failure Five or more internal communications have failed
while "Process Data Update". Last usable value is
displayed
On user error codes: 8, 34, 35, 36, 37, 38, 73, 80,
86, 121, 122, 141
0x10 Bad Sensor Failure On user error codes: 40, 41, 53, 68, 123, 124, 125
0x1C Bad Out of Service Transducer block is in mode "Out of Service"
Channel is not assigned to an AIFB
Channel is not available in running application