9 Supplement
9.1 Technical
data
General data
Parameter, pressure Gauge pressure, absolute pressure, vacuum
Measuring principle Piezoresistive/Thin film DMS
Communication interface None
Materials and weights
Materials, wetted parts
- Process fitting 316Ti
- Diaphragm 316Ti
- Diaphragm with front flush version 316Ti, Hastelloy C4
- Seal, O-ring FPM, FKM, EPDM, NBR
Materials, non-wetted parts
- Internal transmission liquid Synthetic oil, Halocarbon oil
2)3)
- Housing 316Ti
- Terminal housing 316Ti
- Ground terminal 316Ti
- Plug PA
- Cable gland PA, 316Ti
- Plug seal Silicone
- Connection cable PUR
Available cable length max. 40 m
Weight approx.
- Version with plug connector, cable outlet 0.2 kg (0.441 lbs)
- Version with terminal housing 0.35 kg (0.772 lbs)
Torque max. 50 Nm (36.88 lbft)
Output variable
Output signal 4 … 20 mA
Zero and span adjustable via potentiometer ±5 %
Dead time ≤ 1 ms
Step response time (10 … 90 %)
- Standard version ≤ 1 ms
2)
Synthetic
oil: For measuring ranges
up to 16 bar, FDA listed for the food
processing industry. For measuring ranges up to 25 bar not available.
3)
Halocarbon
oil: With version oil
and greasefree, not with vacuum measuring
ranges, not with absolute measuring ranges < 1 bar
abs
.
24 VEGABAR 17
9 Supplement
27636-EN-120531