6 Set up with the display and adjustment module PLICSCOM
VEGABAR 66 • 4 ... 20 mA/HART
36737-EN-130321
Unit
Unitofmeasurement
bar
Temperature unit
°C
3. Activate the selection with [OK]andselect"Units of measure-
ment with [->].
4. Activate the selection with [OK] and select the requested unit with
[->] (in the example mbar).
5. Conrmwith[OK] and move to position correction with [->].
The adjustment unit is thus switched over from bar to mbar.
Proceed as follows to select the temperature unit:
6)
1. Activate the selection with [OK]andselect"Temperature unit
with [->].
2. Activate the selection with [OK] and select the requested unit with
[->] (e.g. °F).
3. Conrmwith[OK].
The temperature unit is hence switched over from °C to °F.
Proceed as follows:
1. Activateinthemenuitem"Position correction"theselectionwith
[OK].
Position correction
Oset
P
=
+0000 mbar
53 mbar
2. Select with [->], e.g. to accept actual measured value.
Position correction
Accept current measured
value?
Accept
Edit
3. Conrmwith[OK]andmovetomin.(zero)adjustmentwith[->].
Proceed as follows:
1. Editthembarvalueinthemenuitem"zero"with[OK].
Zero
000.0 %
P
=
+0000.0 mbar
0000.0 mbar
2. Set the requested mbar value with [+] and [->].
3. Conrmwith[+] and move to span adjustment with [->].
Thezeroadjustmentisnished.
Carry out a position cor-
rection
Carrying out zero adjust-
ment
6)
Selection options: °C, °F.