15
4 Mounting
Chemical seal CSB • for VEGADIF 65
36133-EN-130417
4 Mounting
4.1 Application conditions
Beforemounting,setupandoperation,takenotethatthepressure
transmitter as well as the chemical seal were selected according to
measuringrange,versionandmaterialsuitablefortheprocesscondi-
tions. The load limits must be maintained in order to guarantee the
speciedaccuracy.
Caution:
Indangeroussubstancessuchase.g.oxygen,acetylene,combusti-
bleortoxicproductsaswellasinrefrigeratingplants,compressors,
etc.,thepertinentinstructionsmustbeobservedinadditiontothe
general regulations.
Take note of the following issues in respect to the process and ambi-
ent temperature:
Mountthedierentialpressuretransmitterinsuchawaythatthe
permissible process and ambient temperature limits are neither
underrun nor exceeded.
Taketheinuenceofconvectionandheatradiationintoaccount
Whenselectingthechemicalseals,makesurethatthettingsand
angesarepressureandtemperatureresistant
For this purpose select the suitable material and pressure stage
Mount in such a way that plus and minus side have the same
ambienttemperaturestokeepthetemperatureinuenceslow
4.2 Instructions for handling
Instrumentsmustbeprotectedagainstsoilingandstronguctua-
tions of the ambient temperature
Leave the measuring system in the packaging until mounting to
protect it against mechanical damages
Whenremovingthepackagingandwhenmounting,takespecial
care to avoid mechanical damage and deformation of the dia-
phragm
Do not carry the pressure transmitter by holding the capillary line
Donotbentthecapillarylines.Kinkscancauseleakageandlead
to an increase in response time
Never loosen sealed screws on the chemical seal or the pressure
transmitter
Do not damage the isolating diaphragm: scratches on the isolating
diaphragm(e.g.fromsharpsubjects)arethemainareaswhere
corrosion can occur
4.3 Mounting instructions
Suitable seals must be selected for sealing
Forangemounting,useasealwithasucientlylargeinnerdiam-
eter and place the seal centrically; contact with the diaphragm will
cause measurement deviations
Suitability for the process
conditions
Process and ambient
temperature
Sealing
Seitenansicht 14