6.4 Parameter adjustment - Density
measurement
The instrument is adapted to the application conditions via the pa-
rameter adjustment. The parameter adjustment is carried out with an
adjustment menu.
Information:
Inthisoperatinginstructionsmanual,theinstrument-specicparam-
eters are described. Further general parameters are described in the
operating instructions manual "Display and adjustment module".
Prerequisites
Thefollowingrequirementsmustbefullledforreliableandsafe
operation:
Thetubemustbelled.Theremustbenoairbubblesinthetube
The source container is switched on
A sample point is close to the measuring point
Warning:
When the source container is switched on, the tube must always be
lled.Ifthetubeisempty,therecanbeanincreasedlocaldoserate.
Makesurethatthetubeislledevenincaseofplantdowntime,or
switchthesourcecontainero.
Carry out a setup in the sequence described in the following.
Caution:
Duringtherstsetuporafteraninstrumentresettheinstrumentstarts
with preset standard values. These value are not suitable for your ap-
plication and must be replaced by real values.
Themainmenuisdividedintovesectionswiththefollowingfunc-
tions:
Setup: Settings, e.g. for measurement loop name, isotope, applica-
tion, background radiation, adjustment, signal output
Display: Settings, for example language, measured value display
Diagnosis: Information, for example, of device status, peak value,
simulation
Additional adjustments: Instrument unit, reset, date/time, copying
function
Info: Instrument name, hardware and software version, date of manu-
facture, instrument features
Check if the correct language is already set for the display. If not, you
can change the language in the menu item "Display/Language".
Instrument start
Main menu