Vega MINITRAC 31 Profibus PA Bedienungsanleitung Seite 49

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 96
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 48
49
6 Set up with the display and adjustment module
MINITRAC 31 • Probus PA
43836-EN-131119
Pulse rate of the sensor
Process value
Switching function of the relay
Information:
The simulation is automatically terminated 10 minutes after the last
pressing of a key.
Additional adjustments
In this menu item, the PIN is permanently activated/deactivated. Thus
you protect the sensor data against unauthorized access and unin-
tended changes. The default setting of the PIN is 0000.
This parameter is described in the operating instructions manual
"Display and adjustment module".
In this menu item you can set the actual date and time.
This parameter is described in the operating instructions manual
"Display and adjustment module".
When a reset is carried out, all settings (with only a few exceptions)
are reset. The exceptions are: PIN, language, SIL and HART mode.
The following reset functions are available:
Basic settings: Resetting of the parameter adjustments to default
values at the time of shipment. Order-specic settings are deleted.
Default settings: Resetting of the parameter adjustment like under
"Basic settings". In addition, special parameters are reset to default
values. Order-specic settings are deleted.
Peak values of measured value: Resetting of the parameter adjust-
ments in the menu item "Setup" to the default values of the respective
instrument. Order-specic settings remain but are not taken over into
the current parameters.
Additional settings - PIN
Additional adjustments -
Date Time
Additional adjustments
- Reset
Seitenansicht 48
1 2 ... 44 45 46 47 48 49 50 51 52 53 54 ... 95 96

Kommentare zu diesen Handbüchern