9 Diagnostics and service
FIBERTRAC 31 • Probus PA
43833-EN-131119
Code
Text mes-
sage
Cause Rectication
F036
Faulty pro-
gram memory
Error during software
update
Repeat software update
Exchanging the electronics
F037
RAM hard-
ware error
Error in RAM Restart instrument
Exchanging the electronics
F038
Slave signals
failure
Connection cable to the
Slave instrument interrupted
Instrument not dened as
Slave instrument
Check the connection cable
to the Slave instrument
Dene instrument as Slave
F040
Hardware
error
Sensor defective Restart instrument
Exchanging the electronics
F041
Photomultipli-
er error
Error in the measured value
recording
Restart instrument
Exchanging the electronics
F052
Faulty cong-
uration
Invalid parameter adjust-
ment
Carry out a reset
F066
Faulty adjust-
ment
Adjustment not yet carried
out
Error during adjustment or
when entering the lineariza-
tion table
Carrying out adjustment
Carry out linearization
F068
Count rate too
high
Faulty instrument settings Carry out a reset
F072
Limit exceed-
ed
Faulty instrument settings Carry out a reset
F073
Error real val-
ue correction
Faulty real value correction Carry out real value correc-
tion again
F080
System error
Instrument error Restart instrument
Call our service
F086
Communica-
tion error
Error in the Fieldbus com-
munication
Restart instrument
Call our service
F114
Error real time
clock
Discharge accumulator Readjust real time clock
F120
Filter time
error
Faulty or missing instrument
adjustment
Carrying out adjustment