Vega VEGABAR 40 Profibus PA Bedienungsanleitung Seite 24

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 36
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 23
24 VEGABAR 40 (Profibus PA)
Click in the window "Conditioning“ to Lineari-
sation.
In the window "Span“ you first choose in
which unit the adjustment should be carried
out (mbar, psi, kPa). Enter if you want to edit
the percentage or the current value. If you
want to edit the percentage value, enter the
value (e.g. 80 %). Then click to Save. The
span adjustment is finished.
Sensor parameter adjustment – Lin-
earisation curves
Click in the window "Instrument data param-
eter adjustment“ to Conditioning.
In the window "Zero adjustment“ you choose
in which unit the adjustment should be car-
ried out (mbar, psi, kPa). Then click to Save.
The Zero adjustment is finished.
Setup
Click to Zero or Span to carry out the re-
spective adjustment.
Note:
The vessel or the cable must be unpressu-
rised for the zero adjustment! For the span
adjustment, it should be possible to generate
a defined pressure (e.g. 80 %) in the vessel
or on the cable.
Seitenansicht 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 ... 35 36

Kommentare zu diesen Handbüchern

Keine Kommentare