Instrument modification
6.2 Exchange of electronics
You first have to remove the adjustment mod-
ule, as described in chapter "6.1 Exchange
of adjustment modules“, in order to exchange
the complete electronic module of VEGABAR.
Remove the ground screw and the two
smaller fixing screws connecting the elec-
tronic unit with the housing
Pull the electronic unit to the top and pull
out the plug connection
Proceed in reverse sequence when install-
ing the new electronics module
Note:
No new adjustment is necessary after ex-
changing the electronics module. When
switching on the first time (connection of the
supply voltage) after exchange of the elec-
tronics units, it takes approx. 20 s until the
current measured value is displayed.
Ground
screw
Fixing screw
Fixing screw
Plug
connection
Exchange of the adjustment module
Removal of the adjustment module
Separate VEGABAR from the power
supply
Loosen the screws on the upper side of
the housing and remove the cover or the
display module
Loosen connection cables from the termi-
nals, if necessary pull out the plug con-
nection of the display module
Remove the two screws of the adjustment
module
Remove adjustment module and pull out
plug connection
Insertion of an adjustment module
Plug the plug connection of the new ad-
justment module into the plug-in socket of
the sensor electronics
Fasten the new adjustment module
Reconnect the connection cables, if nec-
essary, also the cable from the display
module
Close cover or display module of VEGA-
BAR