Instrument modification
6.2 Exchange of electronics
You first have to remove the adjustment mod-
ule as described in chapter "6.1 Exchange of
adjustment modules“ to exchange the com-
plete electronics module of VEGABAR.
Then remove the ground screw and the
two smaller fixing screws that hold the
electronics module to the housing
Pull the electronic unit to the top and pull
out the plug connection
Proceed in reverse sequence when install-
ing the new electronics module
Ground
screw
Fixing screw
Fixing screw
Plug
connection
Note:
When switching on the first time (connection
of power supply) after exchange of the elec-
tronics unit, it will take approx. 20 s until the
actual measured value is displayed.