4 Mounting
VEGABAR 51 • 4 … 20 mA/HART
36713-EN-130321
Instrumentsintheversion"Oil and grease free for oxygen"should
be unpacked just before mounting. After the protective cover of the
processttinghasbeenremoved,thelabel"O
2
"ontheprocesstting
is visible.
Danger:
Avoidoil,greaseorcontamination.Explosiondanger!
4.2 Mounting steps
FormountingVEGABAR51,aweldedsocketisrequired.Youcannd
thesecomponentsinthesupplementaryinstructionsmanual"Welded
socket and seals".
Sealthethreadwithresistantsealmaterialontheprocesstting
thread 1½ NPT.
→
Screw VEGABAR 51 into the welded socket. Tighten the hexagon
ontheprocessttingwithasuitablewrench.Wrenchsize,see
chapter"Dimensions".
Warning:
Thehousingmustnotbeusedtoscrewtheinstrumentin!Applying
tightening force can damage internal parts of the housing.
SealtheangeconnectionsaccordingtoDIN/ANSIwithasuitable,
resistant seal and mount VEGABAR 51 with suitable screws.
Usethesealsuitablefortherespectiveprocesstting.Youcannd
thecomponentsinthesupplementaryinstructionsmanual"Welded
socket and seals".
4.3 Mounting steps, tube chemical seal
accordingtoDIN11851
The isolating systems are temperature and pressure-aged at 80 °C
and18bar.Thezeropointisadjustedat22°C±2°Candatorque
of275Nm.Thedenedinstallationpositionis:Tubechemicalseal
horizontal,VEGABAR51vertical.
For mounting proceed as follows:
1. Position VEGABAR 51 with tube chemical seal
2. Tightenthethreadedttingsstep-by-steponceontherightand
on the left
3. HoldVEGABAR51,toavoiddistortionfromthedenedinstalla-
tion position.
Caution:
Apart from mounting, the tube chemical seal must not be permanently
under torsion.
4. Check current after mounting. The current must be between 3.9
and4.1mA.Incaseofadeviatingvalue,losenthreadedtting
and mount again.