42
8Maintenanceandfaultrectication
VEGABAR 51 • Foundation Fieldbus
36715-EN-130321
The serial number is stated on the type label of VEGABAR 51 or on
the delivery note.
8.4 Software update
The software version of VEGABAR 51 can be determined as follows:
on the type label of the electronics
Via the display and adjustment module
via PACTware
You can view all software histories on our website www.vega.com.
Make use of this advantage and get registered for update information
via e-mail.
The following components are required to update the sensor soft-
ware:
Sensor
Voltage supply
VEGACONNECT
PC with PACTware
Currentsensorsoftwareasle
At"www.vega.com/downloads"goto"Software".Selectunder"plics
sensors and instruments","Firmware updates"therespectiveinstru-
mentseriesandsoftwareversion.Loadthezipleviatherightmouse
keywith"Save target as"e.g.onthedesktopofyourPC.Movewith
therightmousekeytothefolderandselect"Extract all".Savethe
extractedles,forexampleonthedesktop.
Connect the signal conditioning instrument to power supply and
provide the connection from the PC to the instrument via the interface
adapter.StartPACTwareandgoviathemenu"Project"totheVEGA
project assistant.Select"USB"and"Set instruments online".Activate
theprojectassistantwith"Start".Theassistantestablishesthecon-
nection automatically and opens the parameter adjustment window
"Sensor # online parameter adjustment".Connectthisparameter
adjustment window before you carry out further steps.
Select with the right mouse key the sensor in the project and go to
"Additional function
".Thenclickto"Software update".Thewindow
"Sensor # software update"opens.PACTwarechecksnowthesensor
data and displays the actual hardware and software version of the
sensor. This takes approximately 60 s.
Pushthebutton"Update software"andselectthepreviouslyextracted
hexle.Thenthesoftwareupdatecanbestarted.Theadditionalles
are installed automatically. Depending on the sensor, this procedure
lastsupto1h.Thenthemessageappears""Software update suc-
cessfully executed".
8.5 Instrument repair
If a repair is necessary, please proceed as follows:
Load sensor software
to PC
Prepar
e update
Load software into sen-
sor
Seitenansicht 41