6 Set up with the display and adjustment module PLICSCOM
VEGABAR 52 • Probus PA - climate compensated
36788-EN-130321
Save value
[->] key to select:
Menu change
Select list entry
Select editing position
[+] key:
Change value of the parameter
[ESC] key:
interrupt input
Jump to next higher menu
The sensor is adjusted via the four keys of the display and adjust-
ment module. The LC display indicates the individual menu items. The
functions of the individual keys are shown in the above illustration.
Approx. 10 minutes after the last pressing of a key, an automatic reset
to measured value indication is triggered. Any values not conrmed
with [OK] will not be saved.
6.4 Setup steps
Before starting the actual parameter adjustment of a Probus PA
sensor, the address setting must rst be carried out. You will nd a
detailed description in the operating instructions manual of the display
and adjustment module or in the online help of PACTware or DTM.
VEGABAR 52 can be used for level as well as for process pressure
measurement. Default setting is level measurement. The mode can be
changed in the adjustment menu.
Depending on the application only the respective subchapter "Level
or process pressure measurement" is of importance. There, you nd
the individual adjustment steps.
Level measurement
Set up VEGABAR 52 in the following sequence:
1. Selecting adjustment unit/density unit
2. Carry out a position correction
3. Carry out min. adjustment
4. Carry out max. adjustment
In the menu item "Adjustment unit" you select the physical unit in
which the adjustment should be carried out, e.g. mbar, bar, psi…
The position correction compensates the inuence of the mounting
position or static pressure on the measurement. It does not inuence
the adjustment values.
Information:
The steps 1, 3 and 4 are not necessary for instruments which are
already preset according to customer specications!
You can nd the data on the type label on the instrument or in the
menu items of the min./max. adjustment.
Adjustment system
A
ddr
ess setting
Level or process pres-
sure measurement
Parameter adjustment
"Level measurement"