5 Connecting to power supply
VEGABAR 54 • 4 … 20 mA/HART
36726-EN-130321
8. Press down the opening levers of the terminals, you will hear the
terminal spring closing
9. Check the hold of the wires in the terminals by lightly pulling on
them
10. Connect the screen to the internal ground terminal, connect the
outer ground terminal to potential equalisation
11. Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
12. Screw the housing cover back on
Theelectricalconnectionishencenished.
Fig. 7: Connection steps 6 and 7
Proceed as follows:
1. Loosen the four screws on the housing base with an Allen key
size4
2. Remove the mounting plate from the housing socket
IP 68 version with exter-
nal housing