9
4 Connecting to power supply
VEGABAR 81 • 4 … 20 mA
46298-EN-131125
7. Check the hold of the wires in the terminals by lightly pulling on
them
8. Connectthescreentotheinternalgroundterminal,connectthe
outer ground terminal to potential equalisation
9. Tighten the compression nut of the cable entry gland. The seal
ring must completely encircle the cable
10. Reinsertthedisplayandadjustmentmodule,ifonewasinstalled
11. Screw the housing cover back on
Theelectricalconnectionisnished.
4.2 Single chamber housing
The following illustration applies to the non-Ex as well as to the Ex-ia
version.
5
1
2
+
( )
(-)
678
4...20mA
3
4
1
Fig.4:Electronicsandterminalcompartment,singlechamberhousing
1 Voltagesupply/Signaloutput
2 For display and adjustment module or interface adapter
3 For external display and adjustment unit or Slave sensor
4 Groundterminalforconnectionofthecablescreen
Electronics and terminal
compartment