56
7 Set up with the display and adjustment module
FIBERTRAC 31 • 4 … 20 mA/HART - four-wire
43386-EN-131119
Dry run protection = the relay will deenergise (safe condition) when
the min. level is reached.
Make sure that you have selected the correct characteristics. See
menu item "Setup/Current output mode".
With this menu item you safeguard the sensor parameters against
unauthorizedorunintentionalmodications.
To avoid possible errors during parameterization in a non-safe user
environment,avericationprocedureisusedthatmakesitpossible
to detect parameterization errors reliably. For this, safety-relevant
parametersmustbeveriedbeforetheyarestoredinthedevice.In
normal operating condition, the instrument is also locked against
parameter changes through unauthorized access.
For this reason, the instrument is shipped in locked conditon. The PIN
in the delivery status is "0000".
Afterachange,allsafety-relevantparametersmustbeveried.For
this purpose, a character string comparison must be carried out. This
is used to check the character respresentation and the communica-
tion channels.
Conrmifthetwocharacterstringsareidentical.Thevericationtexts
are provided in German and with all other menu languages in English.
Inasecondstep,allmodiedsafety-relevantparametersarelisted.
Conrmthemodiedvalues.
If the described process of parameter adjustment was run through
completely and correctly, the instrument will be locked and hence
ready for operation.
Otherwise the instrument remains in the released and hence unsafe
condition.
Lock setup - adjustment