7 Set up with the display and adjustment module
SOLITRAC 31 • 4 … 20 mA/HART - four-wire
43387-EN-131119
points (Delta I). The value is an indication for the reliability and repro-
ducibility of the measurement.
Thehigherthedierencebetweenthetwoadjustmentpoints,
thehigherthedierentialvalue(DeltaI)andthemorereliablethe
measurement. A Delta I value below 10 % is an indication for a critical
measurement.
To increase the Delta I value, you have to increase the distance of the
min. and max. adjustment points in the linearization.
In this menu item you can simulate measured values via the current
output. This allows the signal path to be tested, e.g. through down-
stream indicating instruments or the input card of the control system.
Youcansimulatedierentvalues:
Pulse rate of the sensor
Process value
Current output
Switching function of the relay
Information:
The simulation is terminated automatically 60 minutes after the last
key has been pressed.
Diagnosis/Simulation