3 Product description
SOLITRAC 31 • 4 … 20 mA/HART - four-wire
43387-EN-131119
Modicationstatuselectronicsfrom-07
Theinstrumentisavailableindierentelectronicsversions.Eachver-
sioncanbeidentiedviatheproductcodeonthetypelabel:
StandardelectronicstypePT30E-XX
The scope of delivery encompasses:
Radiation-based sensor
Mounting accessory
Documentation
this operating instructions manual
Safety Manual (SIL) PROTRAC series 30 (Document-ID 42961)
Operating instructions manual "Display and adjustment mod-
ule" (optional)
Ex-specic"Safety instructions" (with Ex versions)
ifnecessary,furthercerticates
3.2 Principle of operation
The instrument is suitable for applications in liquids and bulk solids in
vesselsunderdicultprocessconditions.Thereareapplicationpos-
sibilities in nearly all areas of industry.
The level is detected contactlessly right through the vessel wall.
Neitheraprocessttingnoravesselopeningarerequired.Theinstru-
ment is thus ideal for retro installation.
In radiation-based measurement, a Caesium-137 or Cobalt-60 iso-
tope emits focussed gamma rays that are attenuated when penetrat-
ingthevesselwallandthemedium.ThePVTdetectorontheopposite
side of the tank receives the radiation, whose strength is dependent
on the level. The measuring principle has proven to be very reliable
in conjunction with extreme process conditions because it measures
contactlessly from outside through the vessel wall. The measuring
system ensures maximum safety, reliability and plant availability,
independently of the medium and its properties.
3.3 System limitations
Thereareseveralmeasuringprinciple-specicfactorswhichcaninu-
ence the measuring result. Keep these factors in mind in order to fully
utilize the capabilities of the instrument with respect to measurement
certainty and reproducibility.
The implemented radioactive isotope and its activity must be selected
according to the properties of the vessel and the medium. The neces-
sary radioactive activity must be calculated on the basis of the plant
data.
To this end, make use of our planning service for an optimum layout of
the measurement and selection of the isotope. This applies particu-
larly to SIL applications.
Electronics versions
Scope of delivery
Application area
Functional principle
Activity of the source