Vega VEGABAR 67 4 … 20 mA_HART Bedienungsanleitung Seite 59

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 72
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 58
59
10 Supplement
VEGABAR 67 • 4 ... 20 mA/HART
36742-EN-130321
Process conditions
Thespecicationsofthepressurestageandproducttemperatureareusedasanoverview.The
specicationsonthetypelabelareapplicable.
Process pressure
Ʋ with measuring ranges 0.1 bar
(1.45 psig) or 0.2 bar (2.9 psig)
max. 15 bar (217.6 psig) or max. 20 bar (290 psig)
14)
Ʋ with meas. ranges from 0.4 bar
(5.8 psig)
max. 25 bar (363 psig)
15)
Pressurestage,processtting
Ʋ Threadedtting 316L PN 3, PVDF PN 5
16)
Ʋ Thread 316LPN25,PVDFunpressurized
Ʋ Locktting Unpressurized
Product temperature
Ʋ Suspension cable -12 … +100 °C (+10 … +212 °F)
Ʋ Connection tube -12 … +100 °C (+10 … +212 °F)
Vibration resistance mechanicalvibrationswith4gand5…100Hz
17)
With version with connection tube, vibration resistance 1 g. With lengths > 0.5 m (1.64 ft) the tube
must be supported in addition.
Shock resistance
Acceleration 100 g/6 ms
18)
Electromechanical data - version IP 66/IP 67
Cable entry/plug
19)
Ʋ Single chamber housing 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x
blind plug M20 x 1.5
or:
1 x closing cap ½ NPT, 1 x blind plug ½ NPT
or:
1␣xplug(dependingontheversion),1␣xblindstopper
M20␣x␣1.5
or:
2 x blind plug M20 x 1,5
14)
Limited by the overpressure resistance of the measuring cell.
15)
Limitation by the pressure-tightness of the cable connection.
16)
Limited by the overpressure resistance of the measuring cell.
17)
Tested according to the guidelines of German Lloyd, GL directive 2.
18)
Tested according to EN 60068-2-27.
19)
DependingontheversionM12x1,accordingtoISO4400,Harting,7/8"FF.