6 Set up with the display and adjustment module
MINITRAC 31 • Probus PA
43836-EN-131119
The software addressing is only eective if address 126 or higher is
adjusted on the instrument with the address selection switches.
If a hardware address was adjusted (< 126), then the instrument
signals that the hardware address is active.
Here you can assign a suitable measurement loop name. Push the
"OK" key to start the processing. With the "+" key you change the sign
and with the "->" key you jump to the next position.
You can enter names with max. 19 characters. The character set
comprises:
Capital letters from A … Z
Numbers from 0 … 9
Special characters + - / _ blanks
In this menu item you can adjust the MINITRAC 31 to the integrated
isotope in the source container.
For this purpose, check which isotope is integrated in the source
container. You can nd this information on the type label of the source
container.
With this selection, the sensitivity of the sensor is adapted in an opti-
mum way to the isotope.
The MINITRAC 31 requires this information for the decay compensa-
tion. An annual recalibration is hence not necessary.
Enter the requested parameters via the appropriate keys, save your
settings with [OK] and jump to the next menu item with the [ESC] and
the [->] key.
Enter here, the respective application.
This menu item enables adaptation of the sensor to the requested ap-
plication. You can choose between the following applications: "Level",
"Density", "Limit level", "X-ray alarm" or "Real value correction".
Software addressing
Setup - Measurement
loop name
Setup - Isotope
Setup - Application