10 Supplement
VEGABAR 55 • Foundation Fieldbus
36736-EN-130321
Deviation determined according to the limit point method according to IEC 60770
10)
Applies to the digitalsignaloutput(HART,ProbusPA,FoundationFieldbus)aswellastoana-
logue
current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Deviation with version 0.075 %
Ʋ Turn down 1 : 1 up to 5 : 1 < 0.075 %
Ʋ Turn down > 5 : 1 < 0.015 % x TD
Deviation with version 0.1 %
Ʋ Turn down 1 : 1 up to 5 : 1 < 0.1 %
Ʋ Turn down > 5 : 1 < 0.02 % x TD
Deviation with version 0.2 %
Ʋ Turn down 1 : 1 up to 5 : 1 < 0.2 %
Ʋ Turn down > 5 : 1 < 0.04 % x TD
Deviation with absolute pressure measuring range 0.1 bar
Ʋ Turn down 1 : 1 up to 5 : 1 < 0.25 %
Ʋ Turn down > 5 : 1 < 0.05 % x TD
Inuenceoftheproductorambienttemperature
Thermal change zero signal and output span
Applies to the digital
signaloutput(HART,ProbusPA,FoundationFieldbus)aswellastoana-
logue current output 4 … 20 mA and refers to the set span. Turn down (TD) is the ratio nominal
measuring range/set span.
Thermalchangezerosignalandoutputspan,referencetemperature20°C(68°F):
Ʋ In the compensated temperature
range 0 … +100 °C (+32 … +212 °F)
< (0.05 % + 0.1 % x TD)
Ʋ Outside the compensated tempera-
ture range
< (0.05 % + 0.15 % x TD)
Thermal change, current output
Applies also to the analogue
4 … 20 mA current output and refers to the set span.
Thermal change, current output < 0.05 %/10 K, max. < 0.15 %, each with -40 … +80 °C
(-40 … +176 °F)
0,15 %
-0,15 %
-40°C -20°C
20°C
40°C 60°C 80°C
Fig. 27: Thermal change, current output
10)
Incl. non-linearity, hysteresis and non-repeatability.
Seitenansicht 48