- Raw sensor value, i.e. the
uncalibrated measurement value from the sensor. Unit
derives from 'Sensor_range.unit'
l sensor_range
- "Sensor_range.unit' refers to 'Sensor_value', 'Max/Min_peak_sens
or_value',
'Cal_point_hi/lo'
- includes sensor unit: bar, PSI …; only unit
part of DS-68 is writable
l simulate_primary_value
l simulate_secondary_value_1
l simulate_secondary_value_2
l device status
- "0: ""OK"" 13: ""non-specic error"" 17: ""Cal
span too small"" 34: ""EEPROM memory
fault"" 36: ""ROM memory fault"" 37: ""RAM memory fault"" 40: ""non-specic
hardware fault"" 41: ""Sensor element not found"" 42: ""No leaking pulse"" 43: ""No
trigger signal"" 44: ""EMI error"" 113: ""Communication hardware fault"""
l linearization type
- Possible
types
of linearization are: linear, user dened, cylindrical lying container,
spherical container
- "0: ""Linear"" 1: ""User
def"" 20: ""Cylindrical lying container"" 21: ""Spherical
container"""
l curve_points_1_10
- X and Y values for
the user dened linearization curve
l curve_points_11_20
- X and Y values for
the user dened linearization curve
l curve_points_21_30
- X and Y values for the user dened linearization curve
l curve_points_31_33
- X and Y values for
the user dened linearization curve
l curve status
- Result of table plausibility check
- "0: ""Uninitialized"" 1: ""Good"" 2: ""Not monotonous increasing"" 3: ""Not monotonous
decreasing"" 4: ""Not enough values transmitted"" 5: ""Too many values transmitted""
6: ""Gradient of edge too high"" 7: ""Values not excepted"" 8: ""Table currently
loaded"" 9: ""Sorting and checking table"""
l SUB_DEVICE_NUMBER
l SENSOR_ELEMENT_TYPE
- 0: "non-specic"
l display_source_selector
- Selects
the
type of value that is displayed on the PLICSCOM module
- "0: ""Physical value"" 1: ""P
ercent value"" 2: ""Lin percent value"" 6: ""Out(AI1)"" 7:
""Level"" 8: ""Out(AI2)"" 9: ""Out(AI3)"""
l max_peak_sensor_value
VEGABAR 64 Foun