Pressure transmitter D96/D97 (Profibus PA) 5
Product description
1.2 Types and versions
Housing with
terminals
Cable outlet
axial
Cable outlet lateral Plug connection
lateral
Nominal meas. range (gauge pressure) Gauge press. resistance Low pressure resistance
0…0.1 bar / 0…10 kPa 2.5 bar / 250 kPa -0.1 bar / -10 kPa
0…0.2 bar / 0…20 kPa 5 bar / 500 kPa -0.2 bar / -20 kPa
0…0.4 bar / 0…40 kPa 10 bar / 1 000 kPa -0.4 bar / -40 kPa
0…1.0 bar / 0…100 kPa 25 bar / 2 500 kPa -1.0 bar / -100 kPa
0…2.5 bar / 0…250 kPa 25 bar / 2 500 kPa -1.0 bar / -100 kPa
0…5.0 bar / 0…500 kPa 25 bar / 2 500 kPa -1.0 bar / -100 kPa
0…10.0 bar / 0…1 000 kPa 25 bar / 2 500 kPa -1.0 bar / -100 kPa
0…20.0 bar / 0…2 000 kPa 25 bar / 2 500 kPa -1.0 bar / -100 kPa
Measuring ranges D96
Measuring ranges D97
Nominal meas. range (gauge pressure) Gauge press. resistance Low pressure resistance
0…0.1 bar / 0…10 kPa 15 bar / 1 500 kPa -0.2 bar / -20 kPa
0…0.2 bar / 0…20 kPa 20 bar / 2 000 kPa -0.4 bar / -40 kPa
0…0.4 bar / 0…40 kPa 30 bar / 3 000 kPa -0.8 bar / -80 kPa
0…1.0 bar / 0…100 kPa 35 bar / 3 500 kPa -1.0 bar / -100 kPa
0…2.5 bar / 0…250 kPa 50 bar / 5 000 kPa -1.0 bar / -100 kPa
0…5.0 bar / 0…500 kPa 65 bar / 6 500 kPa -1.0 bar / -100 kPa
0…10.0 bar / 0…1 000 kPa 90 bar / 9 000 kPa -1.0 bar / -100 kPa
0…20.0 bar / 0…2 000 kPa 130 bar / 13 000 kPa -1.0 bar / -100 kPa