6 Set up with the display and adjustment module PLICSCOM
VEGABAR 51 • Probus PA
36714-EN-130321
5. Select the requested unit, e.g. kg/dm³ with [->] and conrm with
[OK], the submenu "Density" appears.
Unit of measurement
Density
0001000
kg/dm³
6. Enter the requested density value with [->] and [+], conrm with
[OK] and move to position correction with [->].
The adjustment unit is thus switched over from bar to m.
Proceed as follows to select the temperature unit:
4)
1. Activate the selection with [OK] and select "Temperature unit
with [->].
2. Activate the selection with [OK] and select the requested unit with
[->] (e.g. °F).
3. Conrm with [OK].
The temperature unit is hence switched over from °C to °F.
Proceed as follows:
1. Activate in the menu item "Position correction" the selection with
[OK].
Position correction
Oset
P
=
+0000 mbar
53 mbar
2. Select with [->], e.g. to accept actual measured value.
Position correction
Accept current measured
value?
Accept
Edit
3. Conrm with [OK] and move to min.(zero) adjustment with [->].
Proceed as follows:
1. Edit the % value in the menu item "Min. adjustment" with [OK].
Min. adjustment
+000.0 %
=
+0000.0 mbar
0000.0 mbar
2. Set the requested percentage value with [+] and [->].
3. Conrm with [OK] and edit the requested mbar value.
4. Set the requested mbar value with [+] and [->].
5. Conrm with [+] and move to max. adjustment with [->].
The min. adjustment is nished.
Carry out a position cor-
rection
C
arry out min. adjustment
4)
Selection options: °C, °F.